FDA Brief

Warning letters Devices

FDA warning letter to

ASP Global, LLC. dba Anatomy Supply Partners, LLC.

High riskIssued Dec 9, 2025Posted Jan 13Center for Devices and Radiological Health

FDA cites ASP Global for quality-system, reporting, and unapproved-change violations involving Safe-T-Fill blood collection devices.

What happened

FDA inspected ASP Global’s Austell, Georgia facility from July 28, 2025, through August 1, 2025, and found the Safe-T-Fill Micro Capillary Collection System devices adulterated and misbranded. The letter cites inadequate supplier controls, missing design-history and 510(k) review, defective medical-device reporting procedures, an unreported correction, unapproved product changes, and labeling that lacks adequate directions for use.

What FDA found

  1. FDA said ASP Global did not adequately implement supplier management procedures for a critical supplier of Safe-T-Fill tubes, including missing supplier evaluations and audits.

    21 CFR 820.50
  2. FDA said the firm did not verify the design history file or whether Safe-T-Fill product changes, such as 300 µL tubes, were covered by the original 510(k) clearances.

    21 CFR 820.30
  3. FDA said the firm’s written MDR procedure did not ensure timely transmission of complete medical device reports.

    21 CFR 803.17(a)(3)
  4. FDA said ASP Global did not report a correction—customer notice and sales hold for Safe-T-Fill tubes used with Magellan LeadCare systems—within 10 working days.

    21 CFR 806.10(a)
  5. FDA said ASP Global marketed significant changes to the cleared Safe-T-Fill devices, including 300 µL, self-sealing, and lithium-heparin bilirubin tubes, without a new 510(k).

    21 U.S.C. § 360(k); 21 CFR 807.81(a)(3)(i)
  6. FDA said labeling for several Micro Capillary Blood Collection tubes lacked adequate directions for use and required in vitro diagnostic information, including the manufacturer’s name and place of business.

    21 U.S.C. § 352(f)(1); 21 CFR 809.10

What FDA wants

FDA asks ASP Global to provide adequate details and documentation for its corrective actions, assess its products against the original clearances and obtain any needed 510(k) clearance before resuming commercialization, and correct its MDR and correction-reporting procedures. The provided text does not state a response deadline.

Who should pay attention

  • laboratories and clinicians using capillary blood collection tubes with lead testing systems
  • distributors and buyers of Safe-T-Fill tubes
  • contract manufacturers and suppliers of IVD blood collection devices

This is a summary. Read the original letter on fda.gov.

ASP Global, LLC. dba Anatomy Supply Partners, LLC.'s FDA history

Recalls in the last 12 months

Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL End Caps for SAFE-T-FILL Blood Gas Capillaries; Fits outer diameter 1.6 - 2.1mm, Model 06 0402 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L; Prepared with Clot Activator and Serum Separator Gel; Red Model/Catalog Number: 07 7120 Product Description: The SAFE-T-FILL Capillary Blood Collection System is…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L; Prepared with Lithium Heparin; Green Model/Catalog Number: 07 6101 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device …
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL MicroHematocrit Capillary Tubes, 50 ¿L; Untreated, Blue Markings Model/Catalog Number: 06 0508 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L; Prepared with Clot Activator; Red Model/Catalog Number: 07 7150 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device use…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 ¿L; Prepared with Sodium Heparin Model/Catalog Number: 06 0009 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 ¿L; Prepared with Balanced Heparin, Model/Catalog Number: 06 0187 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L; Prepared with Sodium Fluoride; Gray Model/Catalog Number: 07 7340 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device u…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L; Prepared with Lithium Heparin and Plasma Separator Gel; Green Model/Catalog Number: 07 7221 Product Description: The SAFE-T-FILL Capillary Blood Collection Syste…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 150 ¿L; Prepared with Sodium Heparin Model/Catalog Number: 06 0007 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
Mixing Fleas for SAFE-T-FILL Blood Gas Capillaries, Model 07 9503
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L; Prepared with Lithium Heparin; Green Model/Catalog Number: 07 7251 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device …
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 150 ¿L; Untreated. Model 06 0199 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L; Prepared with Clot Activator; Red Model/Catalog Number: 07 7151 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device use…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL MicroHematocrit Capillary Tubes, 50 ¿L; Plastic, Ammonium Heparinized, Red Marking Model/Catalog Number: 06 0507 Product Description: A clear, plastic capillary blood collection tube coated with Heparin
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Magnetic Rings for SAFE-T-FILL Blood Gas Capillaries, Model 07 9502
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L; Prepared with Clot Activator; Amber Microtube, Red Model/Catalog Number: 07 7450 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a coll…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L; Prepared with Clot Activator and Serum Separator Gel; Red Model/Catalog Number: 07 7121 Product Description: The SAFE-T-FILL Capillary Blood Collection System is…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 115 ¿L; Prepared with Balanced Heparin Model/Catalog Number: 06 0185 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 115 ¿L; Prepared with Sodium Heparin Model/Catalog Number: 06 0005 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.

Full company page

Similar letters to other companies

  1. Q’s American Best Trading Inc.

    FDA warned Q’s American Best Trading that it lacked required foreign supplier verification for imported cinnamon, corn, and instant noodles.

    High risk
  2. Carive Imports Inc.

    Carive Imports failed to provide required records for imported garlic paste, cookies, noodles and pigeon peas after FDA requested them.

    High risk
  3. Mseafood Corporation

    Mseafood Corporation lacked required foreign supplier verification for imported cocktail sauce, FDA says.

    Medium
  4. Chongqing Qiancan Technology Co., Ltd.

    Chongqing Qiancan Technology, an API maker in China, refused FDA records requests; FDA placed its drugs on import alert 66-79.