High riskIssued Mar 15, 2024Posted Apr 2, 2024Division of Medical Device and Radiological Health Operations Central

FDA warns Beckman Coulter over inadequate quality-system procedures for DxI 9000 immunoassay devices.

What happened

FDA inspected Beckman Coulter's Chaska, MN facility from November 1, 2023, through December 11, 2023. FDA found the DxI 9000 Access Immunoassay analyzer instrument system and assays were adulterated, meaning they did not meet required manufacturing standards, because manufacturing and quality methods did not meet the Quality System regulation (current good manufacturing practice, or CGMP). FDA cited inadequate procedures for risk analysis, corrective and preventive action (CAPA), nonconforming product, and design change.

What FDA found

  1. Risk analysis procedures were inadequate: severity ratings for false negative and false positive results on hepatitis B, troponin I, and digoxin assays were set too low, and occurrence ratings used qualitative estimates even when data allowed semi-quantitative estimates.

    21 CFR 820.30(g)
  2. Corrective and preventive action (CAPA) procedures were inadequate: quality data trending and complaint review did not define when to open a CAPA, and corrective actions were closed without verifying they worked and did not harm the finished device.

    21 CFR 820.100(a)
  3. Nonconforming product procedures were inadequate: procedures did not clearly describe use of the Problem Data Collection Sheet, when a 'no problem found' disposition could be chosen or who approved it, and two procedures defined the same disposition term differently.

    21 CFR 820.90(a)
  4. Design change procedures were inadequate: a DxI 9000 design change was treated as a minor change without completed design inputs, and new verification requirements were not added to design input documents.

    21 CFR 820.30(i)

What FDA wants

FDA tells Beckman Coulter to respond in writing within 15 business days with the specific steps taken, how it will prevent recurrence, documentation of corrections and corrective actions, and a timetable for work not yet complete.

Who should pay attention

  • clinical labs that run DxI 9000 immunoassays
  • hospitals relying on troponin, hepatitis B, and digoxin results
  • IVD makers with similar risk-management and CAPA processes

This is a summary. Read the original letter on fda.gov.

Beckman Coulter Inc.'s FDA history

Recalls in the last 12 months

Oct 28 Class II Beckman Coulter, Inc.
Device
ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Mode…
Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
Oct 28 Class II Beckman Coulter, Inc.
Device
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Mode…
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
Oct 21 Class II Beckman Coulter, Inc.
Device
DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
Oct 21 Class II Beckman Coulter, Inc.
Device
IQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
Oct 14 Class II Beckman Coulter Inc.
Device
DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Part No. C13252
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
Oct 14 Class II Beckman Coulter Inc.
Device
DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
Aug 26 Class II Beckman Coulter Inc.
Device
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Aug 26 Class II Beckman Coulter Inc.
Device
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Mod…
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Aug 12 Class II Beckman Coulter Inc.
Device
REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan Walk…
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
Aug 12 Class II Beckman Coulter Inc.
Device
REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284…
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
Apr 29 Class II Beckman Coulter, Inc.
Device
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
Apr 15 Class II Beckman Coulter, Inc.
Device
MicroScan Neg MIC 3J REF C54814
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
Apr 15 Class II Beckman Coulter, Inc.
Device
MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan pa…
Due to manufacturing issue (incorrectly processing) with in vitro multidrug resistant test.
Apr 1 Class II Beckman Coulter, Inc.
Device
MicroScan Neg MIC 56 REF C42464 UDI-DI code: 15099590731212 MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultativ…
Due to the likely presence of contamination in well(s).
Feb 11 Class II Beckman Coulter, Inc.
Device
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter …
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
Feb 11 Class II Beckman Coulter, Inc.
Device
CellMek SPS Sample Preparation System, REF: C44603, with software
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Jan 14 Class II Beckman Coulter, Inc.
Device
Beckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit
Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
Jan 14 Class II Beckman Coulter, Inc.
Device
Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
Dec 17, 2025 Class II Beckman Coulter, Inc.
Device
Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 81901
It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
Nov 26, 2025 Class II Beckman Coulter Inc.
Device
DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping up to ten fluorescent detection channels us…
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.

Full company page

Similar letters to other companies

  1. Olympus Medical Systems Corporation

    FDA warns Olympus over quality-system failures across six medical device plants, including unvalidated designs and released nonconforming devices.

    High risk
  2. Unilam Co., LTD

    Unilam Co., LTD's UV sunlamps violate quality-system and radiation-safety testing rules, FDA says.

    High risk
  3. Lion Street Medical, d/b/a Pensar Medical LLC

    FDA warns Pensar Medical over unreported wound-therapy injuries and missing device reporting procedures.

    High risk
  4. Koven Technologies, Inc.

    FDA warns Koven Technologies that its Bidop 3 Doppler was labeled for fetal use without clearance and had quality system violations.

    High risk