FDA warning letter to

Medline Industries, LP

High riskIssued Mar 25Posted Apr 7Center for Devices and Radiological Health

Medline's Glen Falls, NY plant made angiographic syringes and kits under flawed CAPA, cleaning and design verification controls, FDA says.

What happened

FDA inspected Medline Industries' NAMIC Division facility at 10 Glens Falls Tech Park, Glen Falls, NY, from December 1 to December 12, 2025. Investigators found the company's corrective and preventive action (CAPA) system did not act on a known syringe-to-manifold disconnection problem even after 221 complaints and 177 medical device reports (MDRs, reports of device-related injury or malfunction sent to FDA), and that cleanroom equipment tops had not been cleaned since November 26, 2023. FDA also found design verification for luer connectors tested only one part without explaining why the others were left out.

What FDA found

  1. The company's own CAPA procedure was not followed: a health hazard review called the syringe disconnection risk low even though its design risk analysis ranked air embolism as the most severe outcome, and corrective actions did not match that risk.

    21 CFR 820.100(a)
  2. When complaint rates stayed above the target set for checking whether the fix worked, the company did not send the CAPA back for more work or open a new one as its procedure requires.

    21 CFR 820.100(a)
  3. Cleanroom cleaning procedures and sign-off forms did not cover the tops of manufacturing equipment, and an investigator saw visible dust on top of four machines.

    21 CFR 820.70(g)(1)
  4. Design verification for luer connectors changed to meet ISO 80369-7 tested only one of many affected connectors and did not document why the others were excluded.

    21 CFR 820.30(f)

What FDA wants

FDA tells Medline to reply in writing within fifteen business days describing the specific steps taken, how it will prevent recurrence, and a timetable for any actions not yet finished. The reply goes by email to CDRHEnforcement@fda.hhs.gov with the subject line "CMS Case 723866".

Who should pay attention

  • hospitals and clinicians who use angiographic procedure kits and manifolds
  • device companies that buy syringes or luer connectors from contract manufacturers
  • quality teams that rely on CAPA effectiveness checks to close complaints

This is a summary. Read the original letter on fda.gov.

Medline Industries, LP's FDA history

  1. Medline Industries, LP

    CGMP/QSR/Medical Devices/PMA/Adulterated/Misbranded

Recalls in the last 12 months

Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64155B.
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as (list truncated)
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: (list truncated)
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: 1) LAPAROTOMY GYN CDS, Medline Kit Number/SKU CDS983529O; 2) GYN ROBOTIC PK-LF, Medline Kit Number/SKU DYNJ52432M; 3) TAH PACK, Medline Kit Number/SKU DYNJ60646A; 4) …
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215P
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: C-SECTION TRAY-LF, Medline Kit Number/SKU OBN471K
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE DRAPE C-SEC FEN CLEAR SCRN WIRE, Medline Product Number/SKU DYNJP6106
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: TUBAL LIGATION RFD, Medline Product Number/SKU DYNJ906291G
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: C-SECTION PACK, Medline Product Number/SKU DYNJ48455A
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: 1) FETAL SURGERY CDS-LF, Medline Kit Number/SKUCDS840214R; 2) MAJOR, Medline Kit Number/SKUDYNJ904851I; 3) ABDOMINAL TRANSPLANT, Medline Kit Number/SKUDYNJ908124B.
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKU DYNJ16127N; 2) NYGH-DELIVERY PACK-LF, Medline Kit Number/SKU DYNJ27206D; 3) OB/GYN, Medline Kit Number/SKU DYNJ9…
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE DRAPE C-SECTION W/POUCH ST 7/CS, Medline Product Number/SKU DYNJP6101
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 21 Class II Medline Industries, LP
Device
Medline surgical convenience kits labeled as: 1) OPEN HEART PACK A & B, Medline Product number/SKU DYNJ04584L; 2) OPEN HEART ACCESSORY PACK, Medline Product number/SKU DYNJT5715A; 3) CABG ACCESSORY PACK, Medline Prod…
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Oct 21 Class II Medline Industries, LP
Device
Medline convenience kit labeled as: FOOT ANKLE. Medline Kit SKU DYNJ65395A
Discrepancy in the shelf-life of a component
Oct 21 Class II Medline Industries, LP
Device
Convenience Kits containing CONMED AirSeal Access Ports ADVANCE LAPAROSCOPIC DYNJ911135A CHN GYN ONCOLOGY DYNJ906731M
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Oct 21 Class II Medline Industries, LP
Device
Medline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Oct 21 Class II Medline Industries, LP
Device
Medline surgical convenience kits labeled as: VEIN HARVESTING PACK, Medline product number/SKU DYNJ51845D
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Oct 21 Class II Medline Industries, LP
Device
Convenience Kits containing CONMED AirSeal Access Ports CHN GYN ONCOLOGY DYNJ906731K DYNJ906731L DYNJ906731N DAVINCI PROSTATECTOMY DYNJG901002A DYNJG901002B DYNJG901002C DYNJG90…
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Oct 21 Class II Medline Industries, LP
Device
Medline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055A; 2) AV SHUNT PACK-LF, Medline Product Number/SKU DYNJ24528M.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.

Full company page

Similar letters to other companies

  1. LGM Pharma Solutions, LLC

    FDA warned LGM Pharma Solutions over manufacturing-quality failures in failing-result investigations and process validation for tablets, including a recalled product.

    High risk
  2. USV Private Limited

    FDA warned USV Private Limited that its sterile injectable manufacturing in Daman, India, had aseptic-processing, facility-design and visible-particulate violations.

    High risk
  3. Next Medical Products Company, LLC

    FDA warns Next Medical Products that its ultrasound gels were contaminated and made without required quality controls or clearance.

    High risk
  4. Conroe Welding Supply, Inc.

    FDA warns Conroe Welding Supply over medical oxygen manufacturing failures, including untested batches and a container with 99% (b)(4).

    High risk