Recalls
When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.
910 recalls in the last 12 months. Newest first.
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jun 10 | Class I | B Braun Medical Inc Device |
Design Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number): SSK H M C 20061890 (560074), SSK S C H (560214), SSK BJC H S (ORIGINAL CUST) (560272), SSK CA P M C (ORIGI… These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection. |
| Jun 10 | Class I | Fresenius Kabi USA, LLC Device |
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004 Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern. |
| Jun 10 | Class I | Fresenius Kabi USA, LLC Device |
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005. Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405671. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | Draeger, Inc. Device |
Atlan A350. Model Number: 8211500. anesthesia workstation The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Whitacre Needle 24 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400868, 405652. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405609. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405658. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in., BD¿Quincke Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405741. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | B Braun Medical Inc Device |
Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333867); 25GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333862); GOVP24BK 24G PENCAN TRAY W/ BUP & DEXTROS (3… These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Quincke Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405621, 405735. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 400866, 405672, 405673, 405674, 405675, 405709, 405721, 40… This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | KAYSERBETTEN GMBH & CO. KG Device |
Pediatric care bed; Product Designation: KayserBett IDA; If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children). |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Numbers: 405637, 405707. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405632. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 10 | Class I | Draeger, Inc. Device |
Atlan A350XL. Model Number: 8621600. anesthesia workstation The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use. |
| Jun 10 | Class I | B Braun Medical Inc Device |
Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90 mm), Bupivacaine 0.75% with Dextrose 8.25% Tray (Kit) (333201); Spinocan 25 Ga. x 3-1/2 in. (90 mm), Bupi… These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection. |
| Jun 10 | Class I | Becton Dickinson & Company Device |
BD¿Spinal Tray with BD¿Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL (marcaine). Catalog Number: 405645. This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing facility inspection, citing specific quality deficiencies related to acceptance testing. |
| Jun 3 | Class I | Wawa Beverage Company Food |
Wawa Fruit Punch Pint (16oz), Plastic bottle Undeclared Milk Allergen, |
| Jun 3 | Class I | Wawa Beverage Company Food |
Wawa Iced Tea Lemon 16 oz Pint, Plastic bottle Undeclared Milk Allergen, |
| Jun 3 | Class I | Legacy Bakehouse LLC Food |
Parmesan & Herb Bagel Crisps, Item Number 18139, Net Weight: 10 lbs. Legacy Bakehouse, LLC, N8W22100 Johnson Dr., Waukesha, WI 53166. Products were made with recalled California Dairies milk powder due to Salmonella. |
| Jun 3 | Class I | Legacy Bakehouse LLC Food |
giant eagle Baked Pita Chips with Parmesan, Garlic & Herb Bagel Crisps, Net Wt 7.33 oz. Shipping Case: 12 ct - 7.33 oz Bags Net Weight: 5.49 lbs. Legacy Bakehouse, LLC, Waukesha WI 53186. Products were made with recalled California Dairies milk powder due to Salmonella. |
| Jun 3 | Class I | Honeyville, Inc Food |
(BULK) SPL-BEV Banana CWT (Banana Smoothie); Super Sack Tote Item# 355-0099 Firm received and used potentially contaminated (Salmonella) ingredient to manufacture various products. |
| Jun 3 | Class I | Wawa Beverage Company Food |
Wawa Diet Lemonade Pint (16oz), Plastic bottle Undeclared Milk Allergen, |
| Jun 3 | Class I | Wawa Beverage Company Food |
Wawa Diet Iced Tea Lemon Pint (16oz),Plastic bottle Undeclared Milk Allergen, |
| Jun 3 | Class I | Legacy Bakehouse LLC Food |
Butter Parsley Bagel Crisps, Item Number 18490, Net Weight: 10 lbs. Legacy Bakehouse, LLC, N8W22100 Johnson Dr., Waukesha, WI 53166. Products were made with recalled California Dairies milk powder due to Salmonella. |
| Jun 3 | Class I | Abiomed, Inc. Device |
Automated Impella Controller (AIC); Product Code: 0042-0000-US; A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures. |
| Jun 3 | Class I | Medline Industries, LP Device |
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) PAIN TRAY, Medline Kit SKU DYNJRA1555; 2) SPINAL BLOCK TRAY WHITACRE 24G, Medline Kit SKU DYNJRA1817; 3) TRAY, 25G WHIT … Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints. |
| Jun 3 | Class I | Bolton Medical Inc. Device |
RelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-34-109-34U 28-N4-34-154-30U 28-N4-34-15… Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system. |
| Jun 3 | Class I | Medline Industries, LP Device |
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) ANESTHESIA TRAY, Medline SKU DYNJRA1355A; 2) SPINAL ANESTHESIA TRAY, Medline SKU DYNJRA2151. Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints. |
| Jun 3 | Class I | Medline Industries, LP Device |
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) SPINAL TRAY W/PENCIL POINT NDL, Medline Kit SKU DYNJRA0245; 2) SPINAL TRAY W/PENCIL POINT NDL, Medline Kit SKU DYNJRA0555; … Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints. |
| Jun 3 | Class I | Medline Industries, LP Device |
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) SPINAL ANESTHESIA TRAY-LF, Medline Kit SKU DYNJRA0102A; 2) SPINAL TRAY-LF, Medline Kit SKU DYNJRA0143B; 3) SPINAL PREP TRAY-… Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints. |
| Jun 3 | Class I | Medline Industries, LP Device |
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: SPINAL BLOCK 22G QUINCKE 5S BU, Medline Kit SKU DYNJRA9026 Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints. |
| Jun 3 | Class I | Abiomed, Inc. Device |
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product… Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console. |
| Jun 3 | Class I | Medline Industries, LP Device |
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) SPINAL BLOCK TRAY, Medline Kit SKU DYNJRA1097; 2) TRAY, SPINAL PENCIL POINT 25G, Medline Kit SKU DYNJRA1154; 3) SPINAL … Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints. |
| Jun 3 | Class I | Medline Industries, LP Device |
Medline kits containing Huons Co. Bupivacaine HCL in Dextrose Injection, USP, labeled as: 1) BLOCK TRAY, Medline Kit SKU DYNJRA1181C; 2) SPINAL BLOCK 25G WHTCRE 5S BU, Medline Kit SKu DYNJRA1860. Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and reported efficacy complaints. |
| May 27 | Class I | Ghirardelli Chocolate Company Food |
GHIRARDELLI PERFECTLY PREMIUM FRAPPE MIX CHOCOLATE FLAVORED NET WT 10lbs (4.54kg) Ingredients: *** nonfat dry milk *** GHIRARDELLI PERFECTLY PREMIUM FRAPPE MIX CHOCOLATE FLAVORED NET WT 30LB (13.61kg) Ingredients: *** n… Potential contamination with Salmonella. |
| May 27 | Class I | Ghirardelli Chocolate Company Food |
GHIRARDELLI PREMIUM HOT COCOA MIX NET WT 32oz (2lbs) 907g Ingredients: *** nonfat dry milk *** Distributed by Ghirardelli Chocolate Company, San Leandro, CA 94578, U.S.A. UPC: 747599620126 Potential contamination with Salmonella. |
| May 27 | Class I | Shining Sea Fish Co. LLC Food |
MA Cohen's Kippered Herring vacuum packed, weight provided at retail, UPC 7232577992 Clostridium botulinum is an uncontrolled hazard |
| May 27 | Class I | Solina U.S. Holding Food |
Solina 6035687 Bayou Blackened Ranch Seasoning, 25lb bag potential to be contaminated with Salmonella |
| May 27 | Class I | PS Seasoning & Spices, Inc. Food |
Non Fat Dry Milk 1#, Part #: 02000014P. Net Wt 1 lb (453g) PS Seasoning & Spices 216 W Pleasant St Iron Ridge, WI 53035. Product was made with milk powder recalled by the supplier due to Salmonella contamination. |
| May 27 | Class I | JOHN B SANFILIPPO & SONS INC Food |
Squirrel Brand Travelers Mix 16 oz can UPC 07223899166, 12 cans per case potential presence of Salmonella |
| May 27 | Class I | PS Seasoning & Spices, Inc. Food |
1P Jalapeno Heat Unit for 300 lbs of Meat, Part #: 180002945. Net Wt 18 lbs (8.20kg). PS Seasoning & Spices 216W Pleasant St Iron Ridge, WI 53035. Product was made with milk powder recalled by the supplier due to Salmonella contamination. |
| May 27 | Class I | Solina U.S. Holding Food |
Solina 6035415 Sourdough Cheese Crouton Seasoning, 50lb bag potential to be contaminated with Salmonella |
| May 27 | Class I | PS Seasoning & Spices, Inc. Food |
1P Original Smokehouse Unit for 300 lbs. of Meat, Part #: 180002964. Net Wt 15.10 lbs (6.85kg). PS Seasoning & Spices 216W Pleasant St Iron Ridge, WI 53035. Product was made with milk powder recalled by the supplier due to Salmonella contamination. |
| May 27 | Class I | Ghirardelli Chocolate Company Food |
GHIRARDELLI PERFECTLY PREMIUM FRAPPE MIX MOCHA NET WT 50oz (3lbs 2oz) 1.42kg Ingredients: *** nonfat dry milk *** Distributed by Ghirardelli Chocolate Company, San Leandro, CA 94578, U.S.A. UPC: 747599662119 Potential contamination with Salmonella. |
| May 27 | Class I | Ghirardelli Chocolate Company Food |
GHIRARDELLI PERFECTLY PREMIUM FRAPPE MIX CLASSIC WHITE NET WT 50oz (3lbs 2oz) 1.42kg Ingredients: *** nonfat dry milk *** GHIRARDELLI PERFECTLY PREMIUM FRAPPE MIX CLASSIC WHITE NET WT 10lbs (4.54kg) Ingredients: *** non… Potential contamination with Salmonella. |
| May 27 | Class I | Fontana Flavors, Inc Food |
C7139D1 Natural Parmesan Cheese Flavor, Net Weight: 50 lbs. bags. Item #: 202405. For Further Processing. Fontana Foods, Janesville, WI. Product was made with milk powder recalled by the supplier due to Salmonella contamination. |
| May 27 | Class I | Fontana Flavors, Inc Food |
C0017D Natural & Artificial White Cheddar Cheese Flavor, Net Weight: 50 lbs. bag. Item #: 8500101. For Further Processing. Fontana Flavors, Janesville, WI. Product was made with milk powder recalled by the supplier due to Salmonella contamination. |
| May 27 | Class I | PS Seasoning & Spices, Inc. Food |
PSW21585 Save A Lot Bind 50#, Part #: 20009709. PS Seasoning & Spices 216 W Pleasant St Iron Ridge, WI 53035. Product was made with milk powder recalled by the supplier due to Salmonella contamination. |