| Nov 19, 2025 | Class I | Seoul Trading, Inc. Food |
Item IL1014A, PIGGY Lotte Crispy Crunch Ice Cream Cone, Strawberry flavor, Keep Frozen, (4/5/160ML), UPC 8801118255045. Product of Korea. The label of recalled product declared: Ingredients: Cone snack, Strawberry syrup… Undeclared peanut. FDA notice H-0191-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230 Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0449-2026 |
| Nov 19, 2025 | Class II | Applied Medical Technology Inc Device |
Applied Medical Technology, Inc., NutraGlide, 5 f - 6F x 55 Nasal Feeding Tube with Stylet and ENfit, REF NFT-05055-I The nasal feeding tube may be occluded FDA notice Z-0494-2026 |
| Nov 19, 2025 | Class II | Philips Medical Systems DMC GmbH Device |
DigitalDiagnost 4 (1) High Performance, (2) Flex/Value; Model Number: (1) 712031, (2) 712032; Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues. FDA notice Z-0464-2026 |
| Nov 19, 2025 | Class II | Stryker Sustainability Solutions Device |
HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation. FDA notice Z-0490-2026 |
| Nov 19, 2025 | Class II | Medline Industries, LP Device |
Medline Renewal LigaSure Maryland Jaw Sealer/Divider Nano Coated, Compatible with FT10 Generator 23 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Rep… Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. FDA notice Z-0475-2026 |
| Nov 19, 2025 | Class I | Siemens Medical Solutions USA, Inc Device |
MAGNETOM Cima.X Upgrade. Model Number: 11689304. There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. FDA notice Z-0431-2026 |
| Nov 19, 2025 | Class II | Medline Industries, LP Device |
Medline Renewal LigaSure Maryland Jaw Sealer/Divider NanoCoated, Compatible w/FT10 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. FDA notice Z-0476-2026 |
| Nov 19, 2025 | Class II | SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION Device |
Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems The ceiling type x-ray tube support CH-200 is used to hold the X-ray tube and collimator to position… Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating shaft. FDA notice Z-0499-2026 |
| Nov 19, 2025 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an … Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton. FDA notice Z-0482-2026 |
| Nov 19, 2025 | Class II | Becton Dickinson & Co. Device |
BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100; The 2D barcodes on these cuvettes contain formatting errors that prevent the analyzing system from correctly reading the expiration date. As a result, the instrument enters an error state and the cuvettes cannot be used. FDA notice Z-0484-2026 |
| Nov 19, 2025 | Class II | Medline Industries, LP Device |
Medline Renewal LigaSure Impact Curved Large Jaw Sealer/Divider, Compatible with FT10 Generator, FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocess… Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. FDA notice Z-0468-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0450-2026 |
| Nov 19, 2025 | Class II | Medline Industries, LP Device |
Medline Renewal LigaSure Exact, Dissector without Nano-Coating 21cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical Cutting And Coagulation Device Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. FDA notice Z-0467-2026 |
| Nov 19, 2025 | Class II | Orthofix U.S. LLC Device |
Virata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The following sizes are affected: 1. POLYAXIAL SCREW, CC, 5.5mm X 30mm; Item Number: TL1-10755030. 2. POLYAXIAL SCREW, CC, 5.5mm … Polyaxial screws do not meet performance standards. FDA notice Z-0478-2026 |
| Nov 19, 2025 | Class II | Miach Orthopaedics Device |
BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000; The field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only. FDA notice Z-0479-2026 |
| Nov 19, 2025 | Class II | NeuroLogica Corporation Device |
GM85 Digital Mobile X-ray imaging System; Model Number: GM85; The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue. FDA notice Z-0480-2026 |
| Nov 19, 2025 | Class II | Sonesta Medical AB Device |
Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table designed for full fluoroscopic imaging of the urinary tract during … Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping. FDA notice Z-0460-2026 |
| Nov 19, 2025 | Class II | GE Healthcare Finland Oy Device |
Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/2… Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment. FDA notice Z-0453-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550 Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0443-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0448-2026 |
| Nov 19, 2025 | Class II | Philips Medical Systems DMC GmbH Device |
CombiDiagnost R90; Model Number: 709031; Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues. FDA notice Z-0462-2026 |
| Nov 19, 2025 | Class II | Siemens Healthcare Diagnostics, Inc. Device |
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6; The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge. FDA notice Z-0491-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0451-2026 |
| Nov 19, 2025 | Class II | GE Healthcare Finland Oy Device |
BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manufacturer Part Number U80296-1R01, revision S Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment. FDA notice Z-0457-2026 |
| Nov 19, 2025 | Class II | Medline Industries, LP Device |
Medline RenewalLigasure Blunt Tip Sealer/Divider, NanoCoated Compatible with FT10 Generator 23 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reproces… Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. FDA notice Z-0472-2026 |
| Nov 19, 2025 | Class II | Abbott Laboratories Device |
Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS, 5.7 and 5.10.1 in the US Product Description: Alinity hq Analyzer Component: … Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results. FDA notice Z-0489-2026 |
| Nov 19, 2025 | Class II | Philips Medical Systems DMC GmbH Device |
ProxiDiagnost N90; Model Number: 706110; Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues. FDA notice Z-0461-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Convenience Kit, BREAST HERNIA DYNJ909125G Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0447-2026 |
| Nov 19, 2025 | Class II | Siemens Healthcare Diagnostics, Inc. Device |
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6… The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge. FDA notice Z-0492-2026 |
| Nov 19, 2025 | Class II | Applied Medical Technology Inc Device |
Applied Medical Technology, Inc., NutraGlide, 5 F- 6 F x 55 Nasal Feeding Tube with Stylet and ENfit, REF NFT-05055-N-I The nasal feeding tube may be occluded FDA notice Z-0495-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0445-2026 |
| Nov 19, 2025 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical Snare Model/Catalog Number: SD-U400U-15 Product Description: These instruments have been desig… Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton. FDA notice Z-0483-2026 |
| Nov 19, 2025 | Class II | Covidien, LP Device |
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT; A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion. FDA notice Z-0485-2026 |
| Nov 19, 2025 | Class II | Applied Medical Technology Inc Device |
Applied Medical Technology, Inc., NutraGlide, 5 F- 6 F x 90 Nasal Feeding Tube with Stylet and ENfit, REF NFT-05090-I The nasal feeding tube may be occluded FDA notice Z-0496-2026 |
| Nov 19, 2025 | Class II | Applied Medical Technology Inc Device |
Applied Medical Technology, Inc., NutraGlide, 6 F- 8 F x 55 Nasal Feeding Tube with Stylet and ENfit, REF NFT-06055-I The nasal feeding tube may be occluded FDA notice Z-0497-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0446-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT AD… Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0442-2026 |
| Nov 19, 2025 | Class II | GE Healthcare Finland Oy Device |
Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment. FDA notice Z-0456-2026 |
| Nov 19, 2025 | Class II | Medline Industries, LP Device |
Medline Renewal Ligasure Blunt Tip Sealer/Divider NanoCoated, Compatible with FT10 Generator 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reproce… Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. FDA notice Z-0473-2026 |
| Nov 19, 2025 | Class II | Abbott Medical Device |
Liberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400 There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC). FDA notice Z-0459-2026 |
| Nov 19, 2025 | Class II | TECHNO-PATH MANUFACTURING LTD. Device |
Multichem ID-B. Model Number: SR103B. Intended for use as an unassayed positive qualitative quality control serum to monitor the precision of laboratory testing procedures for the determination of anti-HIV-1 IgG, anti… Potential of negative recovery of the HBsAg analyte across two (2) lots of Multichem ID-B Positive Control when tested on the Alinity i platform. FDA notice Z-0502-2026 |
| Nov 19, 2025 | Class II | Abbott Medical Device |
Eterna SCS IPG (Implantable Pulse Generator), Model Number 32400 There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Patient Controller (PC). FDA notice Z-0458-2026 |
| Nov 19, 2025 | Class III | Accriva Diagnostics, Inc. Device |
VerifyNow PRUTest Platelet Reactivity Test REF 85225 UDI-DI code: 10711234150078 The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 rece… Due to an device without a premarket clearance being incorrectly package and distributed. FDA notice Z-0500-2026 |
| Nov 19, 2025 | Class II | Siemens Healthcare Diagnostics, Inc. Device |
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4; The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge. FDA notice Z-0493-2026 |
| Nov 19, 2025 | Class II | Philips Respironics, Inc. Device |
SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600 Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini. FDA notice Z-0481-2026 |
| Nov 19, 2025 | Class I | Medline Industries, LP Device |
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment. FDA notice Z-0444-2026 |
| Nov 19, 2025 | Class I | Cerenovus Inc Device |
CEREPAK Detachable Coil System - Name/REF: UNIFORM 2MM X 2.5 CM/FCX100202, UNIFORM 2.5MM X 3.3 CM/FCX100253, UNIFORM 3MM X 5.4 CM/FCX100305, UNIFORM 3.5MM X 6.6 CM/FCX100356, UNIFORM 4MM X 7.5 CM/FCX100407, UNIFORM 4MM X… Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention. FDA notice Z-0434-2026 |
| Nov 19, 2025 | Class II | Medline Industries, LP Device |
Medline Renewal Ligasure Maryland Jaw Sealer/Divider NanoCoated, Compatible with FT10 Generator 44 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Repr… Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged. FDA notice Z-0477-2026 |
| Nov 19, 2025 | Class II | Philips Medical Systems DMC GmbH Device |
Precision CRF; Model Number: 706400; Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues. FDA notice Z-0463-2026 |