Recalls
When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.
5,412 recalls in the last 12 months. Newest first.
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Nov 11 | Class II | BioMed Laboratories, LLC. Drug |
Tychasis Coats Aloe Creme (Allantoin 0.5%), packaged in a) 4oz (NDC 70211-703-04), and b)16 oz (NDC 70211-703-16) jars, Manufactured for: Tychasis Corporation, Dallas, TX 75227. Failed Stability Specifications: OOS stability results included viscosity and appearance results that did not align with the established specifications. |
| Nov 11 | Class II | BioMed Laboratories, LLC. Drug |
Tychasis Coats Aloe Lotion (Allantoin 0.5%) packaged in a)0.5 oz, b)8 oz (NDC 70211-704-08), and c)33 oz (NDC 70211-704-33) bottles, Manufactured for: Tychasis Corporation, Dallas, TX 75227. There is no NDC for the 0.5 o… Failed Stability Specifications: OOS stability results included viscosity and appearance results that did not align with the established specifications. |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
ASCORBIC ACID (PRESERVATIVE FREE) 500MG/ML **50ML**, VMC, Victory Medical Center Pharmacy, Austin Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
ZINC SULFATE (MDV) 10MG/ML **30ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
TESAMORELIN (MDV) 10MG/ML **3ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Fresenius Medical Care Renal Therapies Group, LLC Drug |
DELFLEX Peritoneal Dialysis Solution with 2.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay"safe Exchange Set, 2,500 mL Single dose bag, Sterile and Non-Pyrogenic, Rx only, CAT. NO. 054-25322. Fresenius Medica… Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots. |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
TESAMORELIN/IPAMORELIN (MDV) 5MG/2MG/ML **3ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
SEMAGLUTIDE/L-CARNITINE 5/50MG/ML (1ML MDV), VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
DEXPANTHENOL (MDV) 250MG/ML **30ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
TAURINE (MDV) 50MG/ML **30ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Fresenius Medical Care Renal Therapies Group, LLC Drug |
DELFLEX Peritoneal Dialysis Solution with 4.25% Dextrose, Low Magnesium/ Low Calcium and Attached stay safe Exchange Set, 2,500 mL Single dose bag, Sterile and Non-Pyrogenic, Rx only, CAT.NO. 054-25324. Fresenius Medica… Lack of Assurance of Sterility:complaints have been received related to potential leaks from small perforations in the print area on the primary bag in the affected lots. |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
CJC-1295/IPAMORELIN (MDV) 2/2MG/ML **2ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | American Health Packaging Drug |
Cyanocobalamin Injection, USP, 1,000 mcg/mL, Rx Only, 25 x 1mL -Multiple-Dose Vials per carton, Mfg. by: American Regent, Inc., For Bluepoint Laboratories. Carton NDC 68001-542-60; Vial NDC 68001-542-59 Presence of particulate matter:American Regent, Inc. notified BluePoint Pharmaceuticals Laboratories regarding potential glass particle contamination (foreign glass fragment) identified in the product batch. |
| Nov 11 | Class II | Preferred Pharmaceuticals, Inc. Drug |
Artificial Tears Lubricant Eye Drops, Visine, Pkg Size: 15, Mfg: Geri-Care: Brooklyn, NY. NDC: 68788-7266-0 Lack of Assurance of Sterility |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
TIRZEPATIDE/L-CARNITINE **10/50MG/ML** (MDV) **1ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
CJC-1295/IPAMORELIN (MDV) 2/2MG/ML **5ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
CAGRILINTIDE/GLYCINE/B12 (MDV) 3.75/5/1MG/ML **1ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Victory Medical Center Pharmacy Drug |
THYMOSIN ALPHA-1 (MDV) 3MG/ML **5ML**, VMC, Victory Medical Center Pharmacy, Austin, Texas Lack of Processing Controls |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: PGA Absorbable Surgical Suture; Model/Catalog Number: PSN388V (HSI Item No. 600-8803); Product Description: Sterile, synthetic, absorbable, braided 5-0 violet polyglyco… Products lack 510(k) clearance. |
| Nov 11 | Class II | Illumina, Inc. Device |
Nextseq 550Dx Instrument REF: 20005715, FOR IN VITRO DIAGNOSTIC USE (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission. |
| Nov 11 | Class II | Baxter Healthcare Corporation Device |
TruSystem 7500 surgical table, Model/Catalog Number: 4091000 There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. |
| Nov 11 | Class II | Accriva Diagnostics, Inc. Device |
Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103 Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis. |
| Nov 11 | Class II | GE Medical Systems, LLC Device |
Senographe Pristina 3D GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades). |
| Nov 11 | Class II | Baxter Healthcare Corporation Device |
Mobile column TruSystem 7500 U, Product Code 1730720 The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. |
| Nov 11 | Class I | Cardinal Health 200, LLC Device |
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) STR CATH ANGIO PEDIATRIC PACK , CATALOG #SAN68PAVU3; 2) CATH ANGIO PEDIATRIC PAC… Kits contain infusion sets which were recalled for performance issues. |
| Nov 11 | Class I | Cardinal Health 200, LLC Device |
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: KIT EMERGENCY BASKET GENERAL, CATALOG # RES307732A Kits contain infusion sets which were recalled for performance issues. |
| Nov 11 | Class II | INTELERAD MEDICAL SYSTEMS INCORPORATED Device |
intelerad IntelePACS software A cybersecurity vulnerability has been identified in the software that could allow a malicious user to take over an administrator session, delete or modify image or study data, or lock legitimate users out of the system. Additionally, a software issue has been identified that may result in the unintended deletion of image data from an external archive integration. |
| Nov 11 | Class II | Illumina, Inc. Device |
NovaSeq 6000Dx Instrument REF: 20068232, FOR IN VITRO DIAGNOSTIC USE (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: Silk Nonabsorbable Surgical Suture; Model/Catalog Number: PSN632S (HSI Item No. 600-3506); Product Description: Sterile, natural, nonabsorbable, braided 3-0 black silk … Products lack 510(k) clearance. |
| Nov 11 | Class I | Edwards Lifesciences, LLC Device |
Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Software Version: Not applicable. Product Description: Prosthesis, mitral valve, per… Due to a manufacturing issue which may result in cardiovascular balloon rupturing. |
| Nov 11 | Class II | AVID Medical, Inc. Device |
Halyard MINOR PACK. Model/Catalog Number: VASF105-08. Potential for open seals on header bags, compromising sterility. |
| Nov 11 | Class II | Illumina, Inc. Device |
Illumina DRAGEN Server for NextSeq 550Dx REF: 20086130, (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: Silk Nonabsorbable Surgical Suture; Model/Catalog Number: PSN7772S (HSI Item No. 600-3090); Product Description: Sterile, natural, nonabsorbable, braided 3-0 black silk… Products lack 510(k) clearance. |
| Nov 11 | Class II | Luminex Corporation Device |
LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303 Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Suture; Model/Catalog Number: PSN8696P (Henry Schein Item No. 600-0115); Product Description: Sterile, non-absorbable 7-0 blue polypropylene surgical sutu… Products lack 510(k) clearance. |
| Nov 11 | Class I | Cardinal Health 200, LLC Device |
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: 1) KIT,LABOR KIT, , CATALOG # MA1JV8300C; 2) KIT,LABOR KIT, , CATALOG # MA1J… Kits contain infusion sets which were recalled for performance issues. |
| Nov 11 | Class II | Illumina, Inc. Device |
VeriSeq Onsite Server V2 REF: 20047000 (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission. |
| Nov 11 | Class II | Remel, Inc Device |
Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8660P (HSI Item No. 600-7193); Product Description: Sterile, synthetic, nonabsorbable, monofilamen… Products lack 510(k) clearance. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: PGA Absorbable Surgical Suture; Model/Catalog Number: PSN387V (HSI Item No. 600-9683); Product Description: Sterile, synthetic, absorbable, braided 6-0 violet polyglyco… Products lack 510(k) clearance. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: PGA Absorbable Surgical Suture; Model/Catalog Number: PSN392V (HSI Item No. 600-9613); Product Description: Sterile, synthetic, absorbable, braided 4-0 violet polyglyco… Products lack 510(k) clearance. |
| Nov 11 | Class II | restor3d Inc. Device |
Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery. |
| Nov 11 | Class II | NATUS MEDICAL INCORPORATED Device |
Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011 The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study. |
| Nov 11 | Class II | AVID Medical, Inc. Device |
Halyard ARTHROSCOPY PACK. Model/Catalog Number: SAMM014-21. Shelf label of kit is mislabeled as "Extremity" instead of "Arthroscopy Pack." |
| Nov 11 | Class II | Illumina, Inc. Device |
MiSeq Dx Instrument REF: DX-410-1001 / 15036706, FOR IN VITRO DIAGNOSTIC USE (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission. |
| Nov 11 | Class II | CORIN MEDICAL, LTD. Device |
Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040. Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments. |
| Nov 11 | Class II | GE Medical Systems, LLC Device |
Senographe Pristina GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades). |
| Nov 11 | Class I | Cardinal Health 200, LLC Device |
Cardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump Module Model 8100, labeled as: KIT, CATH LAB ADMIT, CATALOG # ADMMH2133A. Kits contain infusion sets which were recalled for performance issues. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: Silk Nonabsorbable Surgical Suture; Model/Catalog Number: PSN683S (HSI Item No. 600-6411); Product Description: Sterile, natural, nonabsorbable, braided 4-0 black silk … Products lack 510(k) clearance. |
| Nov 11 | Class II | Henry Schein Inc. Device |
Generic Name: Perma Sharp Sutures; Generic Name: PGA Absorbable Surgical Suture; Model/Catalog Number: PSN421V (HSI Item No. 600-6870); Product Description: Sterile, synthetic, absorbable, braided 5-0 undyed polyglyco… Products lack 510(k) clearance. |