When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.

5,412 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Jul 8 Class II Blaine Labs Inc
Drug
TERPENICOL Antifungal Solution, (Undecylenic (10-Undecenoic) acid 25%), 1.0 Fl Oz (29.6 mL), Manufactured By: Blaine Labs, Inc., 11037 Lockport Place, Santa Fe Springs, CA 90670. NDC 63347-600-01
CGMP Deviations; the firm discontinued required stability testing for products on the market still within expiry
Jul 8 Class I Baxter Healthcare Corporation
Device
Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) L&D CONTINUOUS EPIDURAL TRAY, Model Number: PAIN1699A
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) MULTIMED, 7FR, 20CM, 3L, CVC BUNDLE, Model Number: ECVC5995; 2) TRIPLE LUMEN CENTRAL LINE BUNDLE, Model Number: ECVC2625; 3) INTROFLEX 6F, 1L, TB, US, Model Number: STCVC26US; 4) AVA H…
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class I Baxter Healthcare Corporation
Device
Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Jul 8 Class II Siemens Medical Solutions USA, Inc
Device
ARTIS icono ceiling. Model Number: 11328100.
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
Jul 8 Class II Bard Access Systems, Inc.
Device
5605150 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086953 5605240 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 008017410869…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class I Insulet Corporation
Device
Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Jul 8 Class II Hamilton Medical AG
Device
IntelliCuff, Portable automatic cuff pressure controller REF: 951001
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
Jul 8 Class II BD SWITZERLAND SARL
Device
MEDLINE 111 GRAVITY SET W/3 Needle Free Y-Sites REF DYNDTN0555 Catalog Number: SAM2461 UDI code: N/A Engineering Sample/Not for Human Use
Due to engineering sample labeled "Not for Human Use" being inadvertently shipped it the U.S. market.
Jul 8 Class II Medacta Usa Inc
Device
Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10…
Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Pro Wireguided 18-20mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Zimmer Surgical Inc
Device
Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems Dual Hose with CPC Connectors Model/Catalog Number: 60-1812-101-00 Software Version: N/A Product Description: Dual Sterile Extensio…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find another readily available device.
Jul 8 Class I Windstone Medical Packaging, Inc.
Device
A M S, ALIGNED MEDICAL SOLUTIONS, Preop Kit-NS, Pack Number AMS18227, surgical convenience kit
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Jul 8 Class I Windstone Medical Packaging, Inc.
Device
A M S, ALIGNED MEDICAL SOLUTIONS, DSAEK Pack-RX, Pack Number AMS12394B, surgical convenience kit
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Jul 8 Class II Bard Access Systems, Inc.
Device
1174108 PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741154935 1295108 PowerPICC SOLO Catheter 5 Fr Dual-Lumen UDI-DI Code: 00801741034602 1174108D3 PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Wireguided 10-12mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Laborie Medical Technologies Corp
Device
Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI …
Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
Jul 8 Class I Medline Industries, LP
Device
Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagn…
Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) MERCY MEDICAL CIRCUMCISION TRAY, Model Number: CIT6780; 2) CIRCUMCISION TRAY W/1.1CM CIRCLAMP, Model Number: CIT3225A; 3) CIRCUMCISION TRAY, Model Number: CIT6325
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) ARTERIAL LINE KIT, Model Number: ART890B; 2) ULTRASOUND GUIDED IV SECUREMENT SET, Model Number: IVS3350
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Siemens Medical Solutions USA, Inc
Device
ARTIS icono biplane. Model Number: 11327600.
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
Jul 8 Class I Medtronic Heart Valves Division
Device
Medtronic Harmony Delivery Catheter System
Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.
Jul 8 Class II Bard Access Systems, Inc.
Device
4380 ECONOMY SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108563 4382A SOFT TISSUE BIOPSY TRAY UDI-DI Code: 0110885403108501 The Soft Tissue Biopsy devices are indicated for use in the collection of aspirate …
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Pro Wireguided 8-10mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Masimo Corporation
Device
Radius VSM Disposable NIBP Cuff, REF: 4825, 4826
Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) OPHTHALMIC PACK W/LIDO, Model Number: DYNJ57922A
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Accuray Incorporated
Device
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F DYNJ905503G
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class II DEPUY (IRELAND)
Device
Depuy Synthes, ATTUNE Revision Hinge Knee Hinge Housing Components with the below product descriptions and Part Codes. 1. ATTUNE KNEE SYSTEM REVISION HINGE HOUSING COMPONENT AND CROSSPIN B; Part Number: 151730010. …
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Wireguided 18-20mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Bard Access Systems, Inc.
Device
CK000980A PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236150 P4254108 PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741129278 P4254108D PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code:…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Wireguided 6-8mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Bard Access Systems, Inc.
Device
7617408 Groshong NXT ClearVue Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741035258 The Groshong¿ NXT PICC provides short (less than 30 days) or long (greater than 30 days) term peripheral access to the central ve…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class I Baxter Healthcare Corporation
Device
Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Jul 8 Class I Windstone Medical Packaging, Inc.
Device
A M S, ALIGNED MEDICAL SOLUTIONS, Thyroid FNA Pack-RX, Pack Number AMS17729, surgical convenience kit
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Jul 8 Class II Bard Access Systems, Inc.
Device
OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 PIG1260T SAFE-T-CENTESIS TRAY 6FR X 16CM UDI-DI Code: 0110885403108365 TPT1000 THORACENTESIS/PARACENTESIS TRAY UDI-DI Code: 0110885403108488…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II DEPUY (IRELAND)
Device
Depuy Synthes, ATTUNE Revision Hinge Knee Hinge Insert Components with the below product descriptions and Part Codes. 1. ATTUNE KNEE SYSTEM REVISION HINGE INSERT 4 10mm B; Part Number: 151730410. 2. ATTUNE KNEE SY…
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
Jul 8 Class II Siemens Medical Solutions USA, Inc
Device
ARTIS icono floor. Model Number: 11327700.
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
Jul 8 Class II Zimmer Surgical Inc
Device
Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems Single Hose with CPC Connectors Model/Catalog Number: 60-4009-101-00 Software Version: N/A Product Description: Single Sterile Exte…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find another readily available device.
Jul 8 Class I Windstone Medical Packaging, Inc.
Device
A M S, ALIGNED MEDICAL SOLUTIONS, Dr. Lewin Pack, Pack Number AMS7200D, surgical convenience kit
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) BLOCK TRAY, Model Number: DYNJRA1590
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Wireguided 15-18mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Wireguided 12-15mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) LUMBAR PUNCTURE CHOICE, Model Number: DYNDH1215
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class I Insulet Corporation
Device
Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (…
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Jul 8 Class II Baxter Healthcare Corporation
Device
Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables
The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters CABG A CDS CDS982269K CDS982269L CDS982269M CDS982269N CARDIAC MAJOR DYNJ901328D DYNJ901328F HEART A DYNJ903465F DYNJ903465G …
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class II Accuray Incorporated
Device
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.