5,412 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Jul 8 Class II Bard Access Systems, Inc.
Device
CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets provide internal reinforcement to aid in catheter placement.
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Wireguided 8-10mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class I Hamilton Medical AG
Device
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) ARTERIAL ACCESS TRAY W/LIDOCAINE, Model Number: ART1145; 2) Universal Central/Art Line Insertion Kit, Model Number: ART155A; 3) ARTERIAL LINE INSERTION TRAY, Model Number: ART1010B
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Pro Wireguided 6-8mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II W L Gore & Associates, Inc.
Device
GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Region) United States (includes Puerto Rico) and Canada: TSB080806A, TSB081006A, TSB081206A, TSB081506A, TSB081706A, TSB121506A, TSB…
Due to catheter separation
Jul 8 Class II Mentor Texas LP
Device
SYRINGE LUER LOK 10CC-200/BOX BD Product Code: 302995 Mentor Product Code: B9604 Syringe, Piston
A single lot of expired syringes were distributed due to a distribution system control failure.
Jul 8 Class II Covidien LLC
Device
Covidien Shiley Disposable Decannulation Plug REF DDCP UDI-DI code: 10884522004060 Product Codes that include the Covidien Shiley Disposable Decannulation Plug are: Product Number / UDI-DI code 4DCFN 10884522006…
Due to product labeling incorrectly including the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate as the product includes 30 to 35% DEHP.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Pro Wireguided 15-18mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Bard Access Systems, Inc.
Device
1274108D PowerPICC SV Catheter 4 Fr Dual-Lumen REKW2449 00801741087110 3173108D PowerPICC SV Catheter 3 Fr Single-Lumen REKV2227 00801741027536 The PowerPlCC SV catheters are intended for short or long term perip…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Medela Inc
Device
Brand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN Model/Catalog Number: ENF060182LD / 101033086 Software Version: N/A Product Description: Enteral Extension Set, Length 60 Orange-Striped Minibore…
May contain incorrect connectors and will not attach or misconnect to ENFit connections and is thus incompatible. Issue could lead to slight delay in delivery of enteral nutritional fluids or oral medication.
Jul 8 Class II Siemens Medical Solutions USA, Inc
Device
ARTIS icono ceiling. Model Number: 11328100.
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
Jul 8 Class II Boston Scientific Corporation
Device
CRE Pro Wireguided 10-12mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II RAYSEARCH LABORATORIES AB
Device
RayCare 2024A SP4. Oncology Information Systems.
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) PAIN MGMT/ PREF W. MED NDL, Model Number: PAIN0072L; 2) CONTINUOUS EPIDURAL TRAY, Model Number: PAIN1317; 3) CONTINUOUS EPIDURAL TRAY, Model Number: PAIN1579A; 4) CSE TRAY, Model Number: P…
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) DYNJ0621189X, Model Number: SM RADIOLOGY-ANGIO PACK-LF; 2) PAIN1470A, Model Number: CONT. 20G EPI MRI COMPATIBLE; 3) TVS4000L, Model Number: TVS4000 ENDOVENOUS PACK
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class I Baxter Healthcare Corporation
Device
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Jul 8 Class II Bard Access Systems, Inc.
Device
CK000525B PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236129 CK000883 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188800 CK000884A PowerPICC Provena SOLO Catheter 4 Fr…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Bard Access Systems, Inc.
Device
5625240 DuoGlide Short-Term Straight Dialysis Catheter 13 Fr Dual-Lumen UDI-DI Code: 00801741042119 The intended use of the Short-Term Dialysis Catheter is attaining temporary/short-term (less than 30 days) vascula…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Bard Access Systems, Inc.
Device
CK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154003 S4153108BP Provena Midline Catheter 3 Fr Single-Lum…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) SUTURING SET, Model Number: SUT13710
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) PI CVC KIT 3L 7 FR 16 CM, Model Number: ECVC8520; 2) NO CATH CENTRAL LINE KIT, Model Number: CVI4880B; 3) 20G x 6" ARTERIAL CATHETER KIT, Model Number: ART356; 4) MULTIMED 7FR, 16CM, 3L CV…
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Bard Access Systems, Inc.
Device
CK000727 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138478 CK000729 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138713 P4153108D PowerMidline Catheter 3 Fr Single-lumen UDI-DI …
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class I Insulet Corporation
Device
Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529…
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Jul 8 Class II Carl Zeiss Meditec AG
Device
Zeiss Intrabeam Spherical Applicator 1.5 cm REF:304534-6000-557
Spherical applicators (radiation therapy accessory) failed to meet manufacturing specifications.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters OPEN HEART DYNJ909090B OPEN HEART CDS QVH CDS982523M
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) IMAGING BIOPSY TRAY, Model Number: DYNDH1068; 2) BREAST BIOPSY PACK, Model Number: PHS234706I
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) STD, INTRO-FLEX INTRODUCER 6F, C, Model Number: I550BF6C; 2) STD, INTRO-FLEX INTRODUCER 9F, C, Model Number: I505BF9C
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II Bard Access Systems, Inc.
Device
5855240 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107177 5855300 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-D…
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II Masimo Corporation
Device
Radius VSM ECG pre-connected Set, REF: 4842, 4695
Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) SUTURE TRAY, Model Number: SUT12260; 2) CHEST TUBE INSERTION TRAY, Model Number: MNS3390; 3) CONTRACEPTIVE IMPLANT INSERTION KIT, Model Number: MNS9715; 4) 20GX3" RADIAL ART LINE KIT, …
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Jul 8 Class II TELEFLEX LLC
Device
Brand Name: Pilling Wecksorb Product Name: Pilling Wecksorb Neurosponges (Cotton) Pilling¿ Wecksorb Neurosponge Strips (Cotton) Model/Catalog Number: 200100 200101 200102 200103 200104 200105 200106 200108 20…
Due to out-of-specification, or potential out-of-specification, levels of endotoxins
Jul 8 Class II Boston Scientific Corporation
Device
CRE Pro Wireguided 12-15mm 240cm
Potential sterile breach of the pouches in which devices are packaged.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART …
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class II Bard Access Systems, Inc.
Device
CK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI …
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Jul 8 Class II RAYSEARCH LABORATORIES AB
Device
RayCare 2024A SP1. Oncology Information Systems.
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
Jul 8 Class II Medline Industries, LP
Device
Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Jul 8 Class II Remel, Inc
Device
GC Agar Base (150mm) 10/PK R04030; GC Agar Base (100mm) 10/PK R01460
Reduced or no recovery of some strains of Neisseria gonorrhoeae.
Jul 1 Class I George J Howe Company, Inc.
Food
6/11oz Sunflower Seeds Salted
The firm is recalling sunflower seed products due to the presence of an undeclared tree nut allergen, cashew.
Jul 1 Class II Boichik Bagels, Inc.
Food
boichik bagles Tuna Salad Ingredients: albacore tuna, mayonnaise (soybean oil, water, whole eggs and egg yolks, distilled vinegar, salt, sugar, lemon juice concentrate, calcium disodium edta, natural flavors), pickle ju…
Ingredient statement declares eggs, but eggs are omitted from the "contains" allergen statement on the product labels.
Jul 1 Class II Joy Gourmet Foods LLC.
Food
Joy Peach Lassi 16 oz.
Product not adequately pasteurized
Jul 1 Class II Joy Gourmet Foods LLC.
Food
Joy Salty Lassi 16 oz.
Product not adequately pasteurized
Jul 1 Class I George J Howe Company, Inc.
Food
5# Sunflower Seeds C/NS
The firm is recalling sunflower seed products due to the presence of an undeclared tree nut allergen, cashew.
Jul 1 Class I Utz Quality Foods, LLC
Food
2.5oz , 8oz , Zapp's Brand Bayou Blackened Ranch Potato Chips, Plastic Bag.
Potential presence of Salmonella.
Jul 1 Class II Joy Gourmet Foods LLC.
Food
Joy Mint Lassi 16 oz.
Product not adequately pasteurized
Jul 1 Class I Spring & Mulberry Inc.
Food
Spring & Mulberry Pecan Date Date-Sweetened Chocolate, Net Wt. 2.1 oz. (60g), individually packaged in a yellow box, UPC 850055470040, Dist. By Spring & Mulberry Raleigh, NC 27606. Wholesale product packed 48 bars (cadd…
Potential Salmonella contamination.
Jul 1 Class II Boichik Bagels, Inc.
Food
boichik bagels Whitefish Salad Ingredients: Smoked whitefish, mayonnaise (soybean oil, water, whole eggs and egg yolks, distilled vinegar, salt, sugar, lemon juice concentrate, calcium disodium edta, natural flavors), p…
Ingredient statement declares eggs, but eggs are omitted from the "contains" allergen statement on the product labels.
Jul 1 Class I Spring & Mulberry Inc.
Food
1) Spring & Mulberry Pure Dark Date-Sweetened Chocolate, Net Wt. 2.1 oz. (60g), individually packaged in a navy blue box, UPC 850055470002, Dist. By Spring & Mulberry Raleigh, NC 27606. Wholesale product packed 48 bars (…
Potential Salmonella contamination