ReportedClassFirmProduct and reason
May 27 Class II Becton, Dickinson and Company
Device
Swan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
May 20 Class II Dakota Honey Company
Food
Dakota Honey Company Spreadable Spun Honey, Original. Stir before use. Produced in Winner, SD. Packaged in the following sized plastic jars: 1. Net Wt 4 oz (113g) UPC 8 60010-57969 5. 2. Net Wt 12 oz (340g) UPC…
Foreign object (stainless steel dust/flakes or shreds of plastic)
May 20 Class II Ocinet, Inc.
Food
...And Kimchi branded kimchi; packaged in 3 sizes of glass containers (126oz, 63oz, 30oz); 126oz UPC- 8541200408 62oz UPC- 8651200409 30oz UPC- 8541200411
Undeclared allergen ingredient (fish)
May 20 Class II Raja Foods and Vegetables Inc.
Food
Swad Orange Flavored Masala Candy
Contains undeclared FD&C Yellow 6
May 20 Class II Dakota Honey Company
Food
Dakota Honey Company Spreadable Spun Honey, Cinnamon-Infused. Stir before use. Produced in Winner, SD. Packaged in the following sized plastic jars: 1. Net Wt 4 oz (113g) UPC 8 60012-41980 7. 2. Net Wt 12 oz (3…
Foreign object (stainless steel dust/flakes or shreds of plastic)
May 20 Class II Prospector Popcorn, LLC
Food
Prospector Popcorn brand Belgian Chocolate Toffee flavored popcorn; Ingredients: Non-GMO popcorn, brown sugar, corn syrup, butter, vanilla, baking soda, sea salt, white chocolate (sugar, palm kernel fat, skimmed milk pow…
Products contain undeclared soy, specifically soy lecithin.
May 20 Class I Sugar Foods LLC
Food
Kroger HOMESTYLE Cheese AND Garlic CROUTONS NET WT 5 OZ (141g) UPC 0 11110 81353 4 DISTRIBUTED BY THE KROGER CO., CINCINNATI, OHIO 45202
Possible Salmonella in croutons. The firm was notified by their supplier that the non-fat milk powder used to manufacture the product was recalled due to possible salmonella.
May 20 Class II Ferris Coffee and Nut Company, Inc.
Food
Frederik's by meijer Vanilla Bourbon Trail Mix, packed in plastic retail-sized black bag NET WT 9 OZ (255g). On the back of the bag is both lot number and a stamped expiration date. UPC 719283 357212; 12-9oz retail units…
Undeclared allergens (wheat and soy).
May 20 Class I Pure Ground Ingredients, Inc.
Food
Pure Ground Ingredients-Organic Peppermint Leaf Powder. wholesale bulk; tea/food manufacturers. Closed container stored in a cool settings at ambient/room temperature, free of direct sunlight. Packaging: Box with poly li…
Product has the potential to be contaminated with Salmonella.
May 20 Class II Raja Foods and Vegetables Inc.
Food
Swad Mango Flavored Masala Candy
Contains undeclared Blue 1
May 20 Class II Dakota Honey Company
Food
Dakota Honey Company the Honey Flight. 4 Flavors of Spreadable Spun Honey Original - Salted Caramel - Cinnamon - Jalapeno. UPC 8 60010-57968 8. 4 - 4 oz (340g) jars. Net Wt 16 oz (454g). Produced in Winner, SD.
Foreign object (stainless steel dust/flakes or shreds of plastic)
May 20 Class II Dakota Honey Company
Food
Dakota Honey Company Spreadable Spun Honey, Jalapeno. Stir before use. Produced in Winner, SD. Packaged in the following sized plastic jars: 1. Net Wt 4 oz (113g) UPC 8 60012-41982 1. 2. Net Wt 12 oz (340g) UP…
Foreign object (stainless steel dust/flakes or shreds of plastic)
May 20 Class II WE R NUTS
Food
Uncle Giuseppe's Marketplace Milk Chocolate Bridge Mix, 11 oz Net Wt, packed in clear tamper resistant plastic container. UPC 812235023309. Distributed by: WE R NUTS LLC, 99 Seaview Blvd, Port Washington, NY 11050
Contains undeclared allergens (cashews, milk, and soy).
May 20 Class II FINESALER LLC
Food
Alain MILLIAT; MARMELADE ORANGE; 300 g 10.6 oz; packaged in glass jars
Product may contain foreign objects, specifically glass.
May 20 Class II Dakota Honey Company
Food
Dakota Honey Company Spreadable Spun Honey, Salted Caramel. Stir before use. Produced in Winner, SD. Packaged in the following sized plastic jars: 1. Net Wt 4 oz (113g) UPC 8 60012-41981 4. 2. Net Wt 12 oz (340…
Foreign object (stainless steel dust/flakes or shreds of plastic)
May 20 Class II Ascend Laboratories, LLC
Drug
Ascend Laboratories, LLC, Metoprolol Succinate Extended-Release Tablets, USP, 25 mg* Rx Only, Manufactured by: Alkem Laboratories Ltd., India, Distributed by: Ascend Laboratories, LLC, Parsippany, NJ 07054, NDC 67877-590…
Failed Dissolution Specifications
May 20 Class II ENDO USA, Inc.
Drug
Buprenorphine Hydrochloride, Injection, 0.3mg/mL, 5x1 mL Single Dose Vials, Rx only, Manufactured for: Endo USA, Malvern, PA 19855, NDC 42023-179-05.
Crystallization; identified as Buprenorphine free base
May 20 Class II Oasis Medical, Inc.
Drug
Oasis Tears PF, Preservative-Free Lubricant Eye Drops, 10mL/0.34 Fl OZ Bottle, Sterile, Manufactured for: OASIS Medical, Inc., 514 S. Vermont Ave., Glendora, CA 91741, USA, Made in France, NDC 42126-6400-1.
Lack of Assurance of Sterility: The recall is being initiated out of an abundance of caution following FDA observations noted during a recent inspection of Excelvision.
May 20 Class II Stryker Sustainability Solutions
Device
BARD EP XT Steerable, Product Number 200794; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class I Medline Industries, LP
Device
Medline medical convenience kits, containing Namic Manifold, Labeled as: 1. MTS,RT,HANDED/LFT,HRT,KIT, Medline Kit Number/SKU 60010508; 2. DBD-MTS,LEFT HEART,KIT, Medline Kit Number/SKU 60160075; 3. CARDIAC CATH P…
Medline has identified the presence of particulate within the fluid path of the Manifolds.
May 20 Class II Stryker Sustainability Solutions
Device
BARD Dynamic Tip Steerable, Product Number 200131; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Stryker Sustainability Solutions
Device
BARD Dynamic XT Deca Steerable, Product Number 201102, REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class I Medline Industries, LP
Device
Namic Preceptor Manifold, custom, Medline Product Number/SKU 64039307; Extravascular blood pressure transducer
Medline has identified the presence of particulate within the fluid path of the Manifolds.
May 20 Class I Medline Industries, LP
Device
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038402; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Medline has identified the presence of particulate within the fluid path of the Manifolds.
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401908; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II GE Medical Systems, LLC
Device
Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401932; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401923; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Stryker Sustainability Solutions
Device
CS Uni-Directional w Auto ID, Product Number D135303; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II DJO, LLC
Device
Brand Name: DONJOY Product Name: ICEMAN, W/COLDPAD, NS, RH Model/Catalog Number: 11-9099 Software Version: No Product Description: 11-9099 - ICEMAN, W/COLDPAD, NS, RH The IceMan CLASSIC3¿ cold therapy unit helps red…
Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
May 20 Class I Medline Industries, LP
Device
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038200; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Medline has identified the presence of particulate within the fluid path of the Manifolds.
May 20 Class II Jolife AB
Device
LUCAS 2 Model Number: 99576 Catalog Numbers: 99576-000024 and 99576-000011 UDI-DI code: 00883873904565 LUCAS 3 Model Number: 99576 Catalog Number: 99576-000043 UDI-DI code: 00883873861875 LUCAS 3.1 Model Num…
Due to a subset of chest compression systems/devices where the service performed on the systems were not appropriately documented (The documents for the service performed to justify the passing results were not include in the service records).
May 20 Class II Stryker Sustainability Solutions
Device
BARD Dynamic XT Steerable, Product Number 201103, REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401652; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Integra LifeSciences Corp. (NeuroSciences)
Device
Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD Catheter is a 3.3 mm diameter (10 Fr.), 35 cm long polyurethane catheter for diverting cerebrospinal f…
Out-of-specification endotoxin result that did not meet the acceptance criteria.
May 20 Class II GE Medical Systems, LLC
Device
Revolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
May 20 Class II Hologic, Inc
Device
3DIMENSIONS MAMMOGRAPHY SYSTEM; Model Numbers: 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-INTL2D-NS, 3DM-SYS-INTL3D, 3DM-SYS-INTL3D-MOB, 3DM-SYS-INTL3D-NS, 3DM-SYS-STD, 3DM-SYS-STD-MOB, 3DM-SYS-STD-NS.
Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
May 20 Class II DJO, LLC
Device
Brand Name: DONJOY Product Name: ICEMAN W/,UNIV,LOOP,NS,RH Model/Catalog Number: 11-1422 Software Version: no Product Description: 11-1422 - ICEMAN W/,UNIV,LOOP,NS,RH The IceMan CLASSIC3¿ cold therapy unit helps red…
Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401575; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class I Medline Industries, LP
Device
Namic Angiographic Manifold, custom, Medline Product Number/SKU 64038301; Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting.
Medline has identified the presence of particulate within the fluid path of the Manifolds.
May 20 Class II GE Medical Systems, LLC
Device
Revolution Apex, X-ray Computed Tomography, Model Number 5590000-20
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
May 20 Class II Dexcom, Inc.
Device
Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dexcom ONE+ iOS CGM App…
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
May 20 Class II Stryker Sustainability Solutions
Device
BARD Dynamic XT Steerable, Product Number 201104; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Edwards Lifesciences, LLC
Device
Edwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EO
Labeling update to provide warning if functionality of the valve replacement delivery system is compromised.
May 20 Class II Stryker Sustainability Solutions
Device
Medtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Device
Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion…
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
May 20 Class I Medline Industries, LP
Device
1. DBD-MTS,3V,MAN,KIT, Medline Kit Number/SKU 60010195; 2. MTS-LEFT HEART,KIT, Medline Kit Number/SKU 60033123; 3. MTS,3 VALVE,KIT, Medline Kit Number/SKU 60050183; 4. MTS,LEFT HEART,KIT, Medline Kit Number/SKU 600…
Medline has identified the presence of particulate within the fluid path of the Manifolds.
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product
May 20 Class II Stryker Sustainability Solutions
Device
Daig Livewire Steerable, Product Number 401914; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Incomplete seals on sterile product