Recalls
When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.
5,412 recalls in the last 12 months. Newest first.
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Apr 8 | Class II | The Maros Group, LLC Food |
Prickly Pear Jelly. 9 oz (268 g) glass bottle with gold cap. Undeclared milk. FDA notice H-0645-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
STERILE EYE DROPS SOOTHING TEARS (polyethylene glycol 400 0.4% and propylene glycol 0.3%), 0.5 OZ 0.5 fl oz (15 mL) bottles; a) Rugby LUBRICATING EYE DROPS, Distributed by: Rugby Laboratories, Livonia, MI 48152, NDC 053… Lack of Assurance of Sterility FDA notice D-0415-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
Sterile EYE DROPS AC (tetrahydrozoline HCl 0.05%, zinc sulfate 0.25%), 0.5 fl. oz. (15 mL) bottles; a) Best Choice, PROUDLY DISTRIBUTED BY: VALU MERCHANDISERS, CO., 5000 KANSAS AVE, KANSAS CITY, KS 66106, UPC 0 70038 470… Lack of Assurance of Sterility FDA notice D-0409-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
EYE DROPS Advanced Relief, (dextran 70 0.1%, polyethylene glycol 400 1% and tetrahydrozoline HCl 0.05%), 0.5 FL OZ (15 mL) bottles, Sterile; a) Walgreens, DISTRIBUTED BY: WALGREEN CO., DEERFIELD, IL 60015; UPC 3 11917 … Lack of Assurance of Sterility FDA notice D-0410-2026 |
| Apr 8 | Class II | Appco Pharma LLC Drug |
Prazosin Hydrochloride Capsules, USP, 5mg, 100-count bottle, Rx Only, Manufactured for: Biocon Pharma Inc., Iselin, NJ, Manufactured by: Appco Pharma LLC, Piscataway, NJ, NDC 70377-068-11. cGMP deviation: detection of Nitrosamine Drug Substance-Related Impurities, N-nitroso-prazosin impurity C (NNP) above acceptable limits. FDA notice D-0421-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
Sterile EYE DROPS REDNESS LUBRICANT (glycerin 0.25% and naphazoline HCl 0.012%), 0.5 fl oz (15 mL) bottles; a) Industrial Eye Relief, Distributed by: Cintas Corp., Mason, OH 45040, UPC 130209; b) Good Neighbor Pharmacy E… Lack of Assurance of Sterility FDA notice D-0414-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
Dry Eye Relief Eye Drops, (glycerin 0.2%, hypromellose 0.2% and polyethylene glycol 400 1%) 0.5 FL OZ (15 mL) bottles; a) GERI CARE, Distributed by: Gericare Pharmaceuticals Corp., Lakewood, NJ 08701, NDC 57896-181-05; … Lack of Assurance of Sterility FDA notice D-0411-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
Sterile Eye Drops ORIGINAL FORMULA (tetrahydrozoline HCl 0.05%), 0.5 FL OZ (15mL) bottles; a) Good Sense Sterile Eye Drops, Distributed By: Perrigo Direct, Inc., Peachtree City, GA 30269, www.PerrigoDirect.com, NDC 50804… Lack of Assurance of Sterility FDA notice D-0413-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
Ultra Lubricating Eye Drops (polyethylene 400 0.4%, propylene glycol 0.3%), Sterile, 0.5 FL OZ (15mL) bottles; a) LEADER, DISTRIBUTED BY CARDINAL HEALTH, DUBLIN, OHIO 43017, www.myleader.com, NDC 70000-0457-1; b) Harris … Lack of Assurance of Sterility FDA notice D-0412-2026 |
| Apr 8 | Class II | K.C. Pharmaceuticals, Inc Drug |
Artificial Tears Sterile Lubricant Eye Drops (polyvinyl alcohol 0.5%, povidone 0.6%), 0.5 FL OZ (15 mL) bottles; a) LEADER, DISTRIBUTED BY CARDINAL HEALTH, DUBLIN, OHIO 43017, www.myleader.com, NDC 70000-0011-1; b) Good … Lack of Assurance of Sterility FDA notice D-0416-2026 |
| Apr 8 | Class I | Amneal Pharmaceuticals, LLC Drug |
Magnesium Sulfate in Water for Injection, 4g/100 mL (40mg/mL) IV bag, further packaged in cartons of 12 x 100 mL IV bags, Rx only, Manufactured by: Amneal Pharmaceuticals Pvt. Ltd., Ahmedabad, 382110, India, Distributed … Product mix up: a foil pouch labeled "Magnesium Sulfate in Water for Injection, 4 g/100 mL" actually contained an IV bag of Tranexamic Acid instead. FDA notice D-0478-2026 |
| Apr 8 | Class II | Appco Pharma LLC Drug |
Prazosin Hydrochloride Capsules, USP, 2mg, 100-count bottle, Rx Only, Manufactured for: Biocon Pharma Inc., Iselin, NJ, Manufactured by: Appco Pharma LLC, Piscataway, NJ, NDC 70377-067-11. cGMP deviation: detection of Nitrosamine Drug Substance-Related Impurities, N-nitroso-prazosin impurity C (NNP) above acceptable limits. FDA notice D-0420-2026 |
| Apr 8 | Class II | Amneal Pharmaceuticals, LLC Drug |
traMADol Hydrochloride Tablets, USP, 50 mg, 500-count bottles, Rx only, Manufactured by: Amneal Pharmaceuticals Pvt. Ltd., Ahmedabad 382220, India, Distributed by: Amneal Pharmaceuticals LLC, Glasgow, KY 42141, NDC 6021… Failed Impurities/Degradation Specifications: N-nitroso-desmethyl-tramadol impurity (NDSRI) was found out of the specification at 24-month stability interval during Long term stability (25¿C/60%RH). FDA notice D-0408-2026 |
| Apr 8 | Class II | Appco Pharma LLC Drug |
Prazosin Hydrochloride Capsules, USP, 1mg, 100-count bottle, Rx Only, Manufactured for: Biocon Pharma Inc., Iselin, NJ, Manufactured by: Appco Pharma LLC, Piscataway, NJ, NDC 70377-066-11. cGMP deviation: detection of Nitrosamine Drug Substance-Related Impurities, N-nitroso-prazosin impurity C (NNP) above acceptable limits. FDA notice D-0419-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1661-2026 |
| Apr 8 | Class II | Cardinal Health 200, LLC Device |
Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777 The outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the syringes contained within the packages are U-100 Insulin syringes. FDA notice Z-1768-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1739-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1675-2026 |
| Apr 8 | Class II | Synthes (USA) Products LLC Device |
2.4 VOLT(TM) TAP CORTEX SCREW TO 90MM. Part Number: 03.424.126-US. A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread. FDA notice Z-1703-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Allura Xper FD10C; System Code: 722001; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1659-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
MEDLINE ANTERIOR HIP PACK DYNJ64672B Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. FDA notice Z-1711-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1744-2026 |
| Apr 8 | Class II | ReCor Medical Inc. Device |
Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use. FDA notice Z-1654-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Vue Motion V12. Product Number: 1017979. Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence. FDA notice Z-1692-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only); Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1678-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
ALLURA Xper FD20; System Code: (1) 722006, (2)722012, (3)722028; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1663-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, BAT (Breast, Axilla and Thyroid) Coverage. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1758-2026 |
| Apr 8 | Class II | Intuitive Surgical, Inc. Device |
Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument Model/Catalog Number: 420207 Software Version: N/A Product Description: EndoWrist Instruments or Da Vi… Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments FDA notice Z-1700-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Kilt Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1755-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Allura Xper FD20/20; System Code: 722038; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1669-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Allura Xper FD20/15; System Code: 722058; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1671-2026 |
| Apr 8 | Class I | Philips Respironics, Inc. Device |
Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device. In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths. FDA notice Z-1646-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Caps. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1760-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
Medline kits containing Olympus biopsy valves: 1. BAL KIT, DYNDA2137B 2. ENDO KIT, DYKE1915 Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. FDA notice Z-1708-2026 |
| Apr 8 | Class II | Stryker Communications Device |
Stryker CHROMOPHARE Softlit Ring Surgical Light System REF: CH00000001 Surgical light assembly may not adequate support the weight of the ceiling cover. FDA notice Z-1688-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1667-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Thyroid Shield. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1753-2026 |
| Apr 8 | Class II | Civco Medical Instruments Co. Inc. Device |
eTRAX Needle System Starter Kit 12G (for Aurora Trackers), Part Number 667-149 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. FDA notice Z-1731-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1662-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1742-2026 |
| Apr 8 | Class II | Intuitive Surgical, Inc. Device |
Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Ins… Due to increased complaints for broken/frayed grip cables for reusable instruments. FDA notice Z-1697-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
Sterile Radiology Procedure Kits, Model Number DYNDH1491B There is a potential breach in the sterile pouch packaging for one lot of radiology kits, which may compromise product sterility. FDA notice Z-1693-2026 |
| Apr 8 | Class II | SUMMA THERAPEUTICS, LLC Device |
Brand Name: Finesse BTK Multicath Product Name: Injectable Angioplasty Balloon Catheter Model/Catalog Number: FS252251502 Product Description: 5 French sheath, 0.014" guidewire compatible over-the-wire injectable angi… Potential for the balloon in the device to not meet burst specifications. FDA notice Z-1704-2026 |
| Apr 8 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly whi… Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. FDA notice Z-1695-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
Medline convenience kits containing Olympus biopsy valves: 1. BRONCH PACK, DYKE2096 2. NON MB BRONCHOPSCOPY CCH LH PK, DYNJT6182 3. PULMONARY BRONCH PACK, DYNDA2861A Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. FDA notice Z-1706-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Allura Xper FD20/20 OR Table; System Code: 722039; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1670-2026 |
| Apr 8 | Class I | Philips Respironics, Inc. Device |
Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device. In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths. FDA notice Z-1648-2026 |
| Apr 8 | Class I | Baxter Healthcare Corporation Device |
Blue Ventilator Adapter Module, single-patient use circuit, Product Code M07937. Accessory to Baxter Volara Respiratory Therapy system. Urgent Medical Device Correction for the Volara system single-patient use circuit due to the potential that patients or caregivers may be unaware of a possible decrease in oxygen levels (oxygen desaturation), or potential lung tissue injury from overexpansion (barotrauma) when the Volara device is used in the home care environment. This correction is providing the home caregiver specific instructions for use. FDA notice Z-1637-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only); Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1680-2026 |
| Apr 8 | Class II | Civco Medical Instruments Co. Inc. Device |
eTRAX Needle Sensor - 16G(for Aurora Trackers), Part Number 667-158 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. FDA notice Z-1737-2026 |