Recalls
When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.
5,412 recalls in the last 12 months. Newest first.
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Apr 8 | Class II | Intuitive Surgical, Inc. Device |
Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist I… Due to increased complaints for broken/frayed grip cables for reusable instruments. FDA notice Z-1696-2026 |
| Apr 8 | Class II | Mentor Texas, LP. Device |
Artoura Breast Tissue Expanders Reference Numbers: SDC-100UH SDC-110UH SDC-120UH SDC-130UH SDC-135UH SDC-140UH Smooth Ultra High Profile Tissue Expander, 350cc, 455cc, 535cc, 650cc, 700cc, 850cc Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. FDA notice Z-1687-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Blockers. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1764-2026 |
| Apr 8 | Class I | Philips Respironics, Inc. Device |
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device. In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths. FDA notice Z-1649-2026 |
| Apr 8 | Class I | Philips Respironics, Inc. Device |
Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device. In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths. FDA notice Z-1647-2026 |
| Apr 8 | Class II | Mentor Texas, LP. Device |
CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-107MH, SCPX-117MH,SCPX-127MH, SCPX-135MH, SCPX-146MH, SCPX-156MH Smooth Medium Height Tissue Expander, 275cc, 350cc, 450cc, 550cc, 650cc, 800c… Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. FDA notice Z-1683-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1740-2026 |
| Apr 8 | Class II | Reflexion Medical, Inc. Device |
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00 Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose. FDA notice Z-1656-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1677-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1743-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1741-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only); Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1674-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Frontal Aprons Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1756-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 530G Insulin Pump (MMT-551, MMT-751) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1751-2026 |
| Apr 8 | Class II | Ultradent Products, Inc. Device |
Jiffy Original Composite Polisher Cups (Coarse) 12pk, REF 7011. Composite polisher cups may crumble and break apart easily, which could cause a delay in patient treatment. FDA notice Z-1689-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750… Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. FDA notice Z-1710-2026 |
| Apr 8 | Class II | ReCor Medical Inc. Device |
Brand Name: Paradise¿ Ultrasound Renal Denervation System Product Name: Paradise ¿ Catheter Model/Catalog Number: PRDS-068-02 Product Description: Common name: Ablation catheter renal denervation Due to firm distributing a nonconforming product that supposed to be scrapped and not for human use. FDA notice Z-1655-2026 |
| Apr 8 | Class II | Civco Medical Instruments Co. Inc. Device |
eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. FDA notice Z-1735-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1665-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
ENFIT G-TUBE CONNECTOR, MEDLINE Item No. ENFIT1010GC Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process. FDA notice Z-1690-2026 |
| Apr 8 | Class II | Intuitive Surgical, Inc. Device |
Brand Name: da Vinci S, Si Grasping Retractor Instrument Product Name: da Vinci S, Si Grasping Retractor Instrument Model/Catalog Number: 420278 Software Version: NA Product Description: EndoWrist Instruments or Da V… Due to increased complaints for broken/frayed grip cables for reusable instruments. FDA notice Z-1699-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Protective Sleeves. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1759-2026 |
| Apr 8 | Class II | Mentor Texas, LP. Device |
Artoura Breast Tissue Expanders Reference Numbers: SDC-100H SDC-110H SDC-120H SDC-130H SDC-135H SDC-140H SDC-150H SDC-155H Smooth High Profile Tissue Expander, 225cc, 300cc, 375cc, 475cc, 500cc, 600cc, 750cc, 850cc Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. FDA notice Z-1686-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Allura Xper FD20/10; System Code: 722029; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1668-2026 |
| Apr 8 | Class II | Mentor Texas, LP. Device |
CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: SCPX-157LH Smooth Low Height Tissue Expander, 650cc Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. FDA notice Z-1682-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. FDA notice Z-1709-2026 |
| Apr 8 | Class I | Abiomed, Inc. Device |
Purge Cassette, sold within Impella Pump Sets and Individually Packaged. Labeled and distributed in the below configurations and corresponding product codes. 1. Purge Cassette, 5 Pack; Product Code: 0043-0003. 2. … Increased risk of purge leaks with Generation 1 purge cassettes. FDA notice Z-1589-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Kilt Blockers. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1765-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process. FDA notice Z-1691-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1750-2026 |
| Apr 8 | Class II | Mentor Texas, LP. Device |
CPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: 354-9113, 3549313, 354-9314 SILTEX Low Height Tissue Expander Style (9100, 450cc), (9300, 450cc), (9300, 550cc) Infusion sets may contain dull/blunt needle tip, which may be difficult to advance, and/or break. FDA notice Z-1681-2026 |
| Apr 8 | Class II | Siemens Healthcare Diagnostics, Inc. Device |
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536; A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes. FDA notice Z-1657-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
ALLURA Xper FD10F; System Code: 722002; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1660-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Demi Half Aprons. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1763-2026 |
| Apr 8 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLS Model/Catalog Number: TFL-PLS containing TFL-CPLU Product Description: An electricity powered device asse… Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing. FDA notice Z-1694-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1666-2026 |
| Apr 8 | Class I | Abiomed, Inc. Device |
Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0… Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps. FDA notice Z-1591-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1746-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Vest Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1754-2026 |
| Apr 8 | Class II | Burlington Medical, LLC Device |
Burlington Medical, Half Aprons. Potential for attenuation degradation over time, decreasing the lifespan. FDA notice Z-1762-2026 |
| Apr 8 | Class II | Civco Medical Instruments Co. Inc. Device |
eTRAX Needle System Starter Kit 14G (for Aurora Trackers), Part Number 667-150 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. FDA notice Z-1732-2026 |
| Apr 8 | Class II | Oxoid Australia Pty Limited Device |
OXOID MICROBACT IDENTIFICATION KITS, Listeria 12L Kit, 20 Tests; Listeria identification sysystem Oxoid Ltd., part of Thermo Fisher Scientific, confirmed that Microbact 12L Kit 20 Tests MB1128A lot 4494873 may identify incorrect organism due to inconclusive result from ambiguous color reactions in wells. FDA notice Z-1766-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 5 M12; System Code: (1)722227, (2)722231; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1679-2026 |
| Apr 8 | Class II | Civco Medical Instruments Co. Inc. Device |
eTRAX Needle Sensor - 14G(for Aurora Trackers), Part Number 667-157 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. FDA notice Z-1736-2026 |
| Apr 8 | Class II | Medtronic MiniMed, Inc. Device |
Paradigm Insulin Pump (MMT-712, MMT-715) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. FDA notice Z-1748-2026 |
| Apr 8 | Class II | Medline Industries, LP Device |
Airway Exam Kit, DYKE1796 Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. FDA notice Z-1705-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1676-2026 |
| Apr 8 | Class II | Civco Medical Instruments Co. Inc. Device |
eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. FDA notice Z-1733-2026 |
| Apr 8 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. FDA notice Z-1673-2026 |
| Apr 8 | Class II | BD KIESTRA LAB AUTOMATION Device |
BD Kiestra" ReadA; Catalog No.: 446948. In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity. FDA notice Z-1658-2026 |