FDA Brief

Recalls

When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.

5,412 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Resection Sheath, 8 mm. Model No. A42011A. Resection sheath for gynecological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1448-2026
Mar 4 Class II Medline Industries, LP
Device
Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG C…
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
FDA notice Z-1467-2026
Mar 4 Class II Medline Industries, LP
Device
Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
FDA notice Z-1463-2026
Mar 4 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) NAIL KIT, Model Number: DYKM1528
Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.
FDA notice Z-1434-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Resection Sheath, 26 Fr. with Deflecting Obturator; Model No. A22042T. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1444-2026
Mar 4 Class II Medtronic Perfusion Systems
Device
Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
FDA notice Z-1473-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Resection Inner Sheath, for 26 Fr. Outer Sheath. Model No. A22040A. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1439-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Inner Sheath; Model No. A2660. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1447-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Resection Sheath, 28 Fr. with Deflecting Obturator. Model No. A22043T. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1446-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Inner Sheath. Model No. A2641. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1455-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Resection Sheath, 28 Fr.; Model No. A22043A. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1445-2026
Mar 4 Class II Medline Industries, LP
Device
Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medli…
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
FDA notice Z-1465-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Inner Sheath, Long; Model No. WA22017A. Inner sheath for urological and gynecological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1450-2026
Mar 4 Class II MRIMed Inc.
Device
MRI LED Mobile Exam Light Battery REF SL-111 This non-magnetic UL and CE approved battery is used in conjunction with SL-110, Mobile Exam Light
Due to battery component overheating while charging resulting in melting of internal components and causing smoke.
FDA notice Z-1436-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Resection Sheath, 24 Fr., with Deflecting Obturator; Model No. A22041T. Resection sheath for urologic applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1442-2026
Mar 4 Class II Encore Medical, LP
Device
Brand Name: Reverse Shoulder Prosthesis (RSP) Product Name: RSP HUMERAL SOCKET INSERT, 32MM +4MM, STANDARD HXe-plus Model/Catalog Number: 509-00-432
Knee and Humeral socket implants contain incorrect labeling.
FDA notice Z-1458-2026
Mar 4 Class II Olympus Corporation of the Americas
Device
Olympus Sheath, 10 Fr. Model No. A37004A. for endoscopic diagnosis and treatment in urological applications.
Complaints of the ceramic tip of the resection sheath breaking have been received.
FDA notice Z-1454-2026
Mar 4 Class II Medline Industries, LP
Device
Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
FDA notice Z-1468-2026
Mar 4 Class II Medline Industries, LP
Device
Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
FDA notice Z-1464-2026
Feb 25 Class II Woodland Foods LLC
Food
Woodland Gourmet Arrowroot Flour item code H59EE86, Net Wt. 55 lb., bag in box
Possible presence of Foreign Material
FDA notice H-0515-2026
Feb 25 Class II Lotus Mom Corporation dba Indian Kitchen Mart
Food
Brass Tope. Sticker label: $29.99+TAX
Potential contamination with leachable lead.
FDA notice H-0512-2026
Feb 25 Class II Lotus Mom Corporation dba Indian Kitchen Mart
Food
IKM Aluminum Saucepan 9" with Wooden Handle. Sticker label: ALUMINIUM PAN 2 SIDE POUTED WOODEN HANDLE 9" UPC 7 023672 411878 US $ 17.99 Sticker label: IKM INDIAN KITCHEN MART
Potential contamination with leachable lead.
FDA notice H-0511-2026
Feb 25 Class II H & N Group Inc.
Food
Frozen Cooked Medium Shrimp Peeled & Deveined, Tail Off-White Shrimp; 1 lbs retail plastic bag. UPC: 041220333035
Product manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137)
FDA notice H-0516-2026
Feb 25 Class I Reser's Fine Foods, Inc.
Food
The Tuna Salad product container is mislabeled with a Chicken Salad lid, which states: Molly Kitchen's Chicken Salad, 789953, NET WT. 5 LB (2.26 kg) per plastic tub. UPC 7 5810830149 8. Distributed by US Foods, Inc. Rose…
Undeclared fish allergen (tuna). Tuna Salad was mislabeled as Chicken Salad.
FDA notice H-0517-2026
Feb 25 Class II Pure Solutions, Inc.
Food
Pure Factors Professional Nighttime Sleep Formula and Pure Formulas Sleep Reset, Net Wt., 1 fl oz (30 ml) liquid dietary supplement, Glass bottle with dropper cap. (Individual 1 fl oz bottles (not sold in case lots)
Label declares bovine colostrum but does not declare milk allergen.
FDA notice H-0537-2026
Feb 25 Class I Art Monkey LLC dba Why Not Natural
Food
Why Not Natural Moringa Capsules 120 capsule bottle marked with lot # A25G051 on the bottom and with an expiration date of 07/2028 stamped on the bottom.
Potential to be contaminated with Salmonella
FDA notice H-0510-2026
Feb 25 Class II Lotus Mom Corporation dba Indian Kitchen Mart
Food
Aluminum Kadai with Metal Handles Size 5. A-Cook MADE IN INDIA SIZE 5 Sticker label: $$24.99+TX
Potential contamination with leachable lead.
FDA notice H-0513-2026
Feb 25 Class I Lotus Mom Corporation dba Indian Kitchen Mart
Food
Brass Pital Pot 4-quart. Sticker label: BRASS HAMMERED HANDI NO 3 UPC 7 023672 414398 US $ 74.99 Sticker label: IKM INDIAN KITCHEN MART
Potential contamination with leachable lead.
FDA notice H-0514-2026
Feb 25 Class II Olympus Corporation of the Americas
Device
Olympus Thunderbeat 5 mm, 20 cm, Front-Actuated Grip
Firm is initiating a removal due to continued reports of adverse events.
FDA notice Z-1376-2026
Feb 25 Class II Olympus Corporation of the Americas
Device
Olympus Thunderbeat, 5 mm, 45 cm Inline Grip
Firm is initiating a removal due to continued reports of adverse events.
FDA notice Z-1379-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number: DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number: DYNJ56436B; 3) CPNB SETUP TRAY, Model Number: DYNJRA1979; 4) CPNB SETUP TRAY, Model Number: DYNJRA1979A; …
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1386-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1409-2026
Feb 25 Class II Olympus Corporation of the Americas
Device
Olympus Thunderbeat, 5 mm, 35 cm, Inline Grip
Firm is initiating a removal due to continued reports of adverse events.
FDA notice Z-1380-2026
Feb 25 Class II Olympus Corporation of the Americas
Device
Olympus Thunderbeat, 5 mm, 45 cm Front-Actuated Grip
Firm is initiating a removal due to continued reports of adverse events.
FDA notice Z-1374-2026
Feb 25 Class II Olympus Corporation of the Americas
Device
Olympus Thunderbeat, 5 mm, 35 cm, Front-Actuated Grip Type S
Firm is initiating a removal due to continued reports of adverse events.
FDA notice Z-1372-2026
Feb 25 Class II Centinel Spine, Inc.
Device
Brand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
Products were mislabeled as the 6mm product but included the 5 mm product.
FDA notice Z-1426-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326; 2) DRAWER 6-IV FLUIDS GLOVES, Model Number: ACC010326A; 3) DRAWER #5, Model Number: ACC010475; 4) DRAWER 1 NEURO CART, Model Number:…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1399-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) CHEST/BREAST-LF, Model Number: DYNJ83950
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1412-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Surgical Drapes: 1) SHEET,DRAPE,40X58,STERILE, Model Number: DYNJP2410; 2) DRAPE,SHEET,ECONOMY,40X58",ST,20/CS, Model Number: DYNJP2411; 3) SHEET,DRAPE, 40X70,STERILE, Model Number: DYNJP2412; 4) DRAPE,MED…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1423-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTA…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1421-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMU…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1416-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) C-SECTION PACK, Model Number: CDS820080N; 2) C-SECTION CDS, Model Number: CDS983523G; 3) C-SECTION, Model Number: CDSCCR611O; 4) C-SECTION, Model Number: CDSCCR611P; 5) LUD C-SECTION…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1417-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) OR ARTERIOGRAM, Model Number: CDS985094F; 2) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4A; 4) KIT ACES INSERT P…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1389-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840015AQ; 2) OPEN HEART CDS, Model Number: CDS840023T; 3) OFF PUMP CABG CDS, Model Number: CDS840087AI; 4) OFF PUMP CABG CDS, Model Number: CDS840087AK…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1390-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number: PAIN0150C
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1387-2026
Feb 25 Class II Intuitive Surgical, Inc.
Device
Da Vinci 5 ASSY, DV5 CONSOLE, IS5000 Model/Catalog Number: 380730 Used for minimally invasive surgery.
An error event following the release of a software version for the surgical system that can result in loss of user interface content on an external monitor or tower monitor.
FDA notice Z-1428-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ508…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1422-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) HYSTEROSCOPY, Model Number: CDS984086I; 2) KIT GYN ONCOLOGY LAP, Model Number: DYKMBNDL91A; 3) VAGINAL HYSTERECTOMY PACK-LF, Model Number: DYNJ0373807J; 4) VAGINAL HYSTERECTOMY PACK-LF…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1418-2026
Feb 25 Class I Olympus Corporation of the Americas
Device
Olympus High Flow Insufflation Unit. Model Number: UHI. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdomi…
Issue with software algorithm which may lead to overpressure events.
FDA notice Z-1352-2026
Feb 25 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104; 2) KIT OPHTHALMIC RUPTURE GLOBE, Model Number: DYKMBNDL104A; 3) KIT OPHTHALMOLOGY CORNEAL TRAN, Model Number: DYKMBNDL159; 4) K…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA notice Z-1420-2026