Company record
Agiliti Health - Ellis
0 FDA warning letters since 2023 and 1 recall in the last 12 months.
Get an email the day FDA posts anything new about Agiliti Health - Ellis.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Mar 4 | Class II | Agiliti Health - Ellis Device |
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Matt… Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function. FDA notice Z-1472-2026 |