Company record
ARROW INTERNATIONAL, LLC
0 FDA warning letters since 2023 and 47 recalls in the last 12 months.
Get an email the day FDA posts anything new about ARROW INTERNATIONAL, LLC.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT: 1-L 4.5 FR X 15 CM/CDC-41541-JX1A; AGBA PI JACC KIT: 2-L 5.5 FR X 15 CM/CDC-41552-JX1… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION KIT/AK-05502D; EPIDURAL CATHETERIZAT… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 CM/ASK-15123-UOC1, AHDC KIT: 3-L 12 FR X … Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH; PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1; P… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PARACENTESIS KIT/ASK-00376-JHH Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSEL CATH KIT: 20GA X 5IN/AK-04550-S CVC KI… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT: 2L 5.5 FR X 55 CM CG+ VPS/ASK-4555… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-12703-PSPH1; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-WMC2; MULTILUMEN PI CVC KIT: 3L 7 FR X 20CM/ASK-15703-L… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: ACCESS TRAY/ASK-04001-MC3 Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHILIC/ASK-00520-JH; CATH S-L 20 GA X 5 CM H… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: AHDC KIT: 3-L 12 FR X 16 CM/AK-12123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/AK-15123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XP1A; AHDC KIT: 3-L 12 FR X 25… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: SPINAL ANESTHESIA KIT/ASA-24090-SB; SPINAL ANESTHESIA KIT/ASA-24103-SB; SPINAL ANESTHESIA KIT/ASA-25090-SB; SPINAL ANESTHESIA KIT: 24 GA X 9 CM/ASK-24090-SA; SPINAL ANESTHESIA KIT: 25 GA X 9 CM/ASK-24090… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PICC ACCESS TRAY/ASK-04001-CG1 Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: SAC KIT: 18 GA X 12CM W BIO/SAC-01218-B; SAC KIT: 18 GA X 12CM W CHG/SAC-01218-C; SAC KIT: 18 GA X 16CM W BIO/SAC-01618-B Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB CDC-25142-X1A Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MIDLINE 1L: 4.5FR X 15CM/ASK-41541-VFHM; A… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PERITONEAL LAVAGE CATH KIT:8FR/AK-09000 Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-PMAH; 3L PI CVC KIT: 7 FR X 16 CM/… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PACING/PSI KIT: 5 FR 2L/ASK-07155-MGU Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-UPM; RA CATH KIT: 2… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS; STIMUCATH SFTY KIT: 20G CTH, 18Gx8… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: BIOPSY SYSTEM COMPREHENSIVE KIT/ASK-09458-AH, ASK-09458-COC, ASK-09458-EUH, ASK-09458-SD, ASK-09458-VU Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PSI KIT: 7 FR X 4 IN/ASK-09701-UP; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU5; PSI KIT: 8.5 FR/ASK-09807-UC; PSI KIT: 9 FR X 4 1/8IN (10 CM)/ASK-09903-HH2; PSI KIT: 9 FR X 10 CM/ASK-09903-LMDTG; P… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711 Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018-HUMC; ARTERIAL LINE KIT: 18 GA X 6… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-CMC1… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP, DRAINAGE KIT/ASK-01000-NW, THORACENTESI… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
PARACENTESIS KIT, REF: AK-00376, ASK-00376-VCU Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PACING PSI KIT: 6 FR/AI-06210-KS, PACING PSI KIT: 5 FR 2-L/AI-07155-KS, PACING PSI KIT: 5 FR 2L/ASK-07155-UPM Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR x 55CM w Biopatch/DLX-45541-HPKB; PICC/D… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PNHS2; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PNHS2 Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYSIS KIT 2L: 14 FR X 20CM AGB/AK-25142… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| Jun 24 | Class I | ARROW INTERNATIONAL, LLC Device |
Description/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHS1; A… Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
| May 20 | Class I | ARROW INTERNATIONAL, LLC Device |
Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272-VFI, CS-15272-VFIE, CS-15312-VFI, CS-15… 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury. |
| May 20 | Class I | ARROW INTERNATIONAL, LLC Device |
Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury. |
| May 20 | Class I | ARROW INTERNATIONAL, LLC Device |
NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep Antegrade Hemodialysis Catheter with Ar… 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury. |
| May 20 | Class I | ARROW INTERNATIONAL, LLC Device |
Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer,… 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury. |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
MAC" Two-Lumen Central Venous Access Kit for use with 7.5 - 8 Fr. Catheters REF ASK-11142-LC2 UDI code: (01)10801902206807(17)270531(11)251211(10)33F25M0670 The MAC Multi-Lumen Central Venous Access Device permits sh… Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
ARROW RADIAL ARTERY CATHETERIZATION KIT REF ASK-04220-UCL1 UDI code: (01)10801902159127(17)270331(11)251203(10)33F25J0352 REF ASK-04020-EUH2 UDI code: (01)10801902213379(17)270731(11)251218(10)33F25J0869 REF AS… Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
ARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-45703-MGH UDI code: (01)1080190222… Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
FLEXTIP PLUS EPIDURAL CATHETER;FLEXBLOCK CONTINOUS PERIPHERAL NERVE BLOCK CATHETER KIT/SET REF ASK-05502-BSM UDI code: (01)10801902209853(17)270831(11)251210(10)33F25K0409 The Arrow Epidural Catheter kit permits acce… Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
PERCUTANEOUS SHEATH INTRODUCER with 7.5 - 8 Fr. Catheter REF ASK-09903-MGH3 UDI code: (01)10801902159127(17)270331(11)251203(10)33F25J0352 The Arrow¿ percutaneous sheath introducer permits venous access and catheter … Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
Pressure Injectable Arrowg+ard Blue Plus¿ Four-Lumen CVC Kit REF ASK-45854-PUPM1 UDI code: (01)10801902159578(17)270531(11)251205(10)33F25M0011 The Arrowg+ard technology is intended to provide protection against cath… Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
Arrow Echogenic Introducer Needle REF ASK-04001-BWK2 UDI code: (01)10801902220315(17)261231(11)251210(10)33F25J0347 This device is intended for injection or aspiration of fluids. The needle protection device covers … Due to receiving a notice from a supplier that the liquid adhesive were incorrectly manufactured |
| Apr 29 | Class II | ARROW INTERNATIONAL, LLC Device |
Arrowg+ard Blue¿ MAC" Two-Lumen Central Venous Access Kit for use with 7.5 - 8 Fr. Catheters REF ASK-21142-DMC UDI code: (01)10801902218930(17)270630(11)251211(10)33F25M0671 The MAC Multi-Lumen Central Venous Acces… Due to receiving a notice from a supplier that the liquid adhesive was incorrectly manufactured |