FDA Brief

Company record

Auris Health, Inc

0 FDA warning letters since 2023 and 2 recalls in the last 12 months.

Get an email the day FDA posts anything new about Auris Health, Inc.

Recalls

ReportedClassFirmProduct and reason
Feb 25 Class II Auris Health, Inc
Device
MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
FDA notice Z-1383-2026
Nov 26, 2025 Class II Auris Health, Inc
Device
MONARCH Bronchoscope. Model Number: MBR-000211-B
Potential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks.
FDA notice Z-0543-2026