Company record
Beckman Coulter Mishima K.K.
0 FDA warning letters since 2023 and 3 recalls in the last 12 months.
Get an email the day FDA posts anything new about Beckman Coulter Mishima K.K..
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Oct 7 | Class II | Beckman Coulter Mishima K.K. Device |
Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer meas… Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results. |
| Mar 18 | Class II | Beckman Coulter Mishima K.K. Device |
CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281 A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment |
| Mar 18 | Class II | Beckman Coulter Mishima K.K. Device |
DxC 700 AU, REF: B86444, B86446 A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment. |