Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Nov 4 | Class II | Boston Scientific Neuromodulation Corporation Device |
Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals. |
| Aug 12 | Class I | Boston Scientific Neuromodulation Corporation Device |
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems. Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement. |
| Jul 15 | Class II | Boston Scientific Neuromodulation Corporation Device |
Intracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESS INSTRUMENTS- DRILL EU/RLV0035-EU, INTRACEPT ACCESS INSTRUMENTS- 2VB/RLV0045, INTRACEPT ACCESS INSTRUMENTS- … Intraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility breach, and if breached could lead to severe infection. |
| Jun 17 | Class II | Boston Scientific Neuromodulation Corporation Device |
Boston Scientific CSK Electrodes: REF: CSK-TC10, CSK Electrode, STAINLESS STEEL, 10 cm x 28 ga (0.36 mm) REF: CSK-TC10-3M,CSK Electrode, STAINLESS STEEL, 3M Cable, 10 cm x 28 ga (0.36 mm); REF: CSK-TC15, CSK Electrode… Reusable electrodes may not meet expected performance levels. |
| Jun 17 | Class II | Boston Scientific Neuromodulation Corporation Device |
Boston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180 Sterile anchors my lack sterility assurance. |
| Jun 17 | Class II | Boston Scientific Neuromodulation Corporation Device |
Boston Scientific TCN Electrodes: REF: TCN-5, TCN Electrode, NITINOL, 5 cm x 0.45 mm (0.0175 in); REF:TCN-10, TCN Electrode, NITINOL, 10 cm x 0.45 mm (0.0175 in) REF:TCN-15, TCN Electrode, NITINOL, 15 cm x 0.45 mm (0.… Reusable electrodes may not meet expected performance levels. |