Recalls

ReportedClassFirmProduct and reason
Nov 4 Class II Canon Medical System, USA, INC.
Device
ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000F
Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
Nov 4 Class II Canon Medical System, USA, INC.
Device
ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000C
Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
Nov 4 Class II Canon Medical System, USA, INC.
Device
ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V
Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
Oct 14 Class II Canon Medical System, USA, INC.
Device
Alphenix INFX-8000C/B5
Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
Sep 2 Class II Canon Medical System, USA, INC.
Device
Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagno…
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Feb 4 Class II Canon Medical System, USA, INC.
Device
Alphenix INFX-8000F, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Feb 4 Class II Canon Medical System, USA, INC.
Device
Alphenix INFX-8000H, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Feb 4 Class II Canon Medical System, USA, INC.
Device
Alphenix INFX-8000V, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Feb 4 Class II Canon Medical System, USA, INC.
Device
Alphenix INFX-8000C, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Jan 21 Class II Canon Medical System, USA, INC.
Device
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Jan 21 Class II Canon Medical System, USA, INC.
Device
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.