| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis MedBank Tower REF: 1145-00 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis Mini Main REF: 349 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis MedStation ES REF: 323 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis Anesthesia Station ES REF: 327 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis MedStation 4000 REF: 303 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis MedBank Mini REF: 1147-00 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Jun 10 | Class II | CareFusion 303, Inc. Device |
BD Pyxis MedStation ES Tower REF: 352 Medication cabinet Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Apr 1 | Class II | CareFusion 303, Inc. Device |
BD Alaris" System with Guardrails" Suite MX with Point of Care Unit Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units |
| Feb 4 | Class II | CareFusion 303, Inc. Device |
BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of flu… Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network |
| Jan 14 | Class II | CareFusion 303, Inc. Device |
BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136451-01, 136451-02, 1… Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems. |
| Dec 31, 2025 | Class II | CareFusion 303, Inc. Device |
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with… Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs. |
| Nov 26, 2025 | Class I | CareFusion 303, Inc. Device |
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Insp… If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated. |