Recalls

ReportedClassFirmProduct and reason
Oct 21 Class II Cook Incorporated
Device
Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
May 13 Class I Cook Incorporated
Device
Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic procedures are available in a variety of French sizes, endhole sizes, lengths, materials and designs (e.g., po…
Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
May 13 Class I Cook Incorporated
Device
Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and …
Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
May 13 Class I Cook Incorporated
Device
Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G31214 NR5.0-35-100-P-10S-PIG-CSC-20 G31216 NR5.0-35-70-P-10S-VCF-CSC-20 G31220 NR5.0-35-100-P-10S-VCF…
Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Arterial Pressure Monitoring Tray, Reference Part Number APMY-2.5-15-0-2.5 , Order Number G58412; Reference Part Number APMY-4.0-21J-1-12, Order Number G58419
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set: Reference Part Number C-TCCSB-500-UNI, Order Number G32193; Reference Part Number C-TCCSB-500-UNI-01, Order Number G47665.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Cook¿ Strange Bile Duct Stone Exploration Set: Reference Part Number C-CSRC-7.5D-40-STRANGE, Order Number G09108
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Wayne Pneumothorax Set: Reference Part Number C-UTPT-1020-WAYNE-IMH, Order Number G56532; Reference Part Number C-UTPT-1400-WAYNE-112497-IMH, Order Number G56535.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Blue Rhino¿ G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-HC-G-NA, Order Number G57706; Refere…
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Approach CTO-18 Micro Wire Guide, Reference Part Number CMW-14-190-18G, Order Number G50788; Reference Part Number CMW-14-300-18G, Order Number G50792
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Pa…
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Spectrum¿ Central Venous Catheter Set: Reference Part Number C-UDLM-401J-ABRM-HC, Order Number G43983; Reference Part Number C-UDLM-401J-ABRM-HC-CAH, Order Number G24754; Reference Part Number C-UDLM-40…
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Approach CTO-6 Micro Wire Guide, Reference Part Number CMW-14-190-6G, Order Number G50786
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Thal-Quick Chest Tube Tray: Reference Part Number C-TQTSY-3200, Order Number G07242
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-HC-G-EU, Order Number G57685; Reference Part Number C-PTIS-100-HC-G-NA, Order Number G57682; Reference Part…
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Multipurpose Drainage Catheter: Reference Part Number ULT8.5-38-25-P-CLMS-8-RH-FS, Order Number G29775
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Ciaglia Blue Rhino Percutaneous Tracheostomy Introducer Set: Reference Part Number C-PTIS-100-FLEX-HC, Order Number G32092; Reference Part Number C-PTIS-100-HC, Order Number G12115.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL NCompass¿ Nitinol Stone Extractor: Reference Part Number C-NTSE-2.4-115-NC3, Order Number G36250; Reference Part Number C-NTSE-2.4-115-NCT4, Order Number G36251.
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Approach CTO-12 Micro Wire Guide: Reference Part Number CMW-14-190-12G, Order Number G50787; Reference Part Number CMW-14-300-12G, Order Number G50791
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Wayne Pneumothorax Tray: Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY-1400-WAYNE-112497-IMH, Order Number G56537; Reference Part Number C-UTPTY…
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Cook¿ Spectrum¿ Central Venous Catheter Tray: Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0003, Order Number G56424; Reference Part Number C-UTLMY-701J-ABRM-CUSTOM-0018, Order Number G21053; Referenc…
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Apr 15 Class II Cook Incorporated
Device
COOK MEDICAL Cook Staged Extubation Set: Reference Part Number C-CAE-14.0-83-EXTU, Order Number G24073
Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.
Dec 31, 2025 Class II Cook Incorporated
Device
Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9.0-18/38-45-RB-ANL1-HC; 3) KCFW-10.0-35-45-RB-HFANL0-HC; 4) KCFW-10.0-35-45-RB-HFANL1-HC; 5) KCFW-12.0-3…
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
Dec 31, 2025 Class II Cook Incorporated
Device
R¿sch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The R¿sch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
Dec 31, 2025 Class II Cook Incorporated
Device
Ring Transjugular Intrahepatic Access Sets, Reference Part Numbers: 1) RTPS-100; 2) RTPS-100-10.0; The Ring Transjugular Intrahepatic Access Set is intended for transjugular liver access in diagnostic and interven…
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.