Company record
CooperSurgical, Inc.
1 FDA warning letter since 2023 and 2 recalls in the last 12 months.
Get an email the day FDA posts anything new about CooperSurgical, Inc..
Warning letters
-
DrugsCooperSurgical, Inc.
FDA says CooperSurgical’s Paragard video omitted all risk information, making the birth-control promotion misleading and misbranded.
Low
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 30 | Class I | CooperSurgical, Inc. Device |
INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715. Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored. |
| Sep 30 | Class I | CooperSurgical, Inc. Device |
INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715. Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored. |