Recalls

ReportedClassFirmProduct and reason
Jul 22 Class II Daavlin Distributing Company
Device
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Jul 22 Class II Daavlin Distributing Company
Device
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Jul 22 Class II Daavlin Distributing Company
Device
Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Jun 10 Class II Daavlin Distributing Company
Device
Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physician, Model Numbers 21…
Some of the Calibration/Output Certificates had the outputs reversed.