Company record
Daavlin Distributing Company
0 FDA warning letters since 2023 and 4 recalls in the last 12 months.
Get an email the day FDA posts anything new about Daavlin Distributing Company.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Jul 22 | Class II | Daavlin Distributing Company Device |
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). |
| Jul 22 | Class II | Daavlin Distributing Company Device |
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). |
| Jul 22 | Class II | Daavlin Distributing Company Device |
Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). |
| Jun 10 | Class II | Daavlin Distributing Company Device |
Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physician, Model Numbers 21… Some of the Calibration/Output Certificates had the outputs reversed. |