FDA Brief

Company record

Datascope Corp.

0 FDA warning letters since 2023 and 7 recalls in the last 12 months.

Get an email the day FDA posts anything new about Datascope Corp..

Recalls

ReportedClassFirmProduct and reason
Mar 25 Class III Datascope Corp.
Device
Battery Charging Station; Model: 0998-00-0802;
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
FDA notice Z-1587-2026
Mar 11 Class II Datascope Corp.
Device
CS300 IABP. Software Version CS300 IABP C.01.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
FDA notice Z-1487-2026
Mar 11 Class II Datascope Corp.
Device
CS100 IABP. Software Version CS100 IABP Q.01.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.
FDA notice Z-1486-2026
Dec 24, 2025 Class III Datascope Corp.
Device
Cardiosave Rescue
The IFU addendum updates the Vibration and Shock Table to reference the correct standards.
FDA notice Z-0932-2026
Dec 24, 2025 Class III Datascope Corp.
Device
Cardiosave Hybrid
The IFU addendum updates the Vibration and Shock Table to reference the correct standards.
FDA notice Z-0931-2026
Dec 17, 2025 Class II Datascope Corp.
Device
Cardiosave Rescue. Intra-Aortic Balloon Pump system.
The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.
FDA notice Z-0917-2026
Dec 17, 2025 Class II Datascope Corp.
Device
Cardiosave Hybrid. Intra-Aortic Balloon Pump system.
The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.