Recalls

ReportedClassFirmProduct and reason
Aug 19 Class II Erbe USA Inc
Device
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Apr 1 Class I Erbe USA Inc
Device
Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
Probes may rupture/burst during activation
Apr 1 Class I Erbe USA Inc
Device
Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use
Probes may rupture/burst during activation
Apr 1 Class I Erbe USA Inc
Device
Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
Probes may rupture/burst during activation