Recalls

ReportedClassFirmProduct and reason
Oct 21 Class II Ferno-Washington Inc
Device
Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No. 0015816 and PowerX2 Premium, Item No. 0015821
Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.
Jul 29 Class II Ferno-Washington Inc
Device
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. …
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.