Company record
Foundation Medicine, Inc.
0 FDA warning letters since 2023 and 5 recalls in the last 12 months.
Get an email the day FDA posts anything new about Foundation Medicine, Inc..
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Aug 26 | Class II | Foundation Medicine, Inc. Device |
Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx. Software issue that can cause an increase in spontaneous software closures. |
| Apr 29 | Class II | Foundation Medicine, Inc. Device |
Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990, 01994, 01995. Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR. |
| Apr 29 | Class II | Foundation Medicine, Inc. Device |
Sequencing Agent SEQ0067, component of FoundationOne CDx (F1CDx) Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants |
| Dec 3, 2025 | Class II | Foundation Medicine, Inc. Device |
FoundationOne CDx (F1CDx), RAL-0003 version 31.0 Reports were sent to customers without the companion diagnostic (CDx) Claims Page. Claims pages were distributed via amended reports within 7 days of the issue occurrence. |
| Dec 3, 2025 | Class II | Foundation Medicine, Inc. Device |
FoundationOne Liquid CDx (F1LCDx), TECH-0009 version 6.0 Reports were sent to customers without the companion diagnostic (CDx) Claims Page. Claims pages were distributed via amended reports within 7 days of the issue occurrence. |