GE Medical Systems Information Technologies Inc
0 FDA warning letters since 2023 and 14 recalls in the last 12 months.
Get an email the day FDA posts anything new about GE Medical Systems Information Technologies Inc.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 23 | Class II | GE Medical Systems Information Technologies Inc Device |
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events. |
| Sep 2 | Class II | GE Medical Systems Information Technologies Inc Device |
PORTRAIT CORE SERVICES SOFTWARE V1.0.5, Ref# 5697353 Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area. |
| Sep 2 | Class II | GE Medical Systems Information Technologies Inc Device |
PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592 Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area. |
| Sep 2 | Class II | GE Medical Systems Information Technologies Inc Device |
FRU (Field replaceable unit), PORTRAIT CORE SERVICES SW RELOAD V1.1, Ref# 5700225 Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area. |
| Sep 2 | Class II | GE Medical Systems Information Technologies Inc Device |
PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209 Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area. |
| Aug 12 | Class II | GE Medical Systems Information Technologies Inc Device |
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient. |
| Aug 12 | Class II | GE Medical Systems Information Technologies Inc Device |
GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information (PHI). |
| Jun 10 | Class II | GE Medical Systems Information Technologies Inc Device |
CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3 Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury. |
| Jun 3 | Class II | GE Medical Systems Information Technologies Inc Device |
GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(Wit… GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). |
| Jun 3 | Class II | GE Medical Systems Information Technologies Inc Device |
GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). |
| Jun 3 | Class II | GE Medical Systems Information Technologies Inc Device |
GE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia … GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). |
| Jun 3 | Class II | GE Medical Systems Information Technologies Inc Device |
GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). |
| Jun 3 | Class II | GE Medical Systems Information Technologies Inc Device |
GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia De… GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). |
| Jun 3 | Class II | GE Medical Systems Information Technologies Inc Device |
GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection… GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). |