FDA Brief

Company record

GE Medical Systems SCS

0 FDA warning letters since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about GE Medical Systems SCS.

Recalls

ReportedClassFirmProduct and reason
Jul 1 Class II GE Medical Systems SCS
Device
GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System
GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.