Company record
IMRIS Imaging Inc
0 FDA warning letters since 2023 and 3 recalls in the last 12 months.
Get an email the day FDA posts anything new about IMRIS Imaging Inc.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 2 | Class II | IMRIS Imaging Inc Device |
ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000 The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure. |
| Sep 2 | Class II | IMRIS Imaging Inc Device |
ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600 The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure. |
| Jan 14 | Class II | IMRIS Imaging Inc Device |
Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000, 119630-000, 113803-000, 113803-600, 113802-000, 113802-600 Software Version: Not Applicable Product Descripti… Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury. |