Company record
Innovasis, Inc
0 FDA warning letters since 2023 and 1 recall in the last 12 months.
Get an email the day FDA posts anything new about Innovasis, Inc.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Aug 12 | Class II | Innovasis, Inc Device |
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step. |