FDA Brief

Company record

Inpeco S.A.

0 FDA warning letters since 2023 and 2 recalls in the last 12 months.

Get an email the day FDA posts anything new about Inpeco S.A..

Recalls

ReportedClassFirmProduct and reason
Dec 3, 2025 Class II Inpeco S.A.
Device
FlexLab (FLX); Version: FLX-217-40;
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
FDA notice Z-0596-2026
Dec 3, 2025 Class II Inpeco S.A.
Device
FlexLab (FLX); Version: FLX-217-10;
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
FDA notice Z-0595-2026