FDA Brief

Company record

Ion Beam Applications S.A.

0 FDA warning letters since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about Ion Beam Applications S.A..

Recalls

ReportedClassFirmProduct and reason
Mar 18 Class II Ion Beam Applications S.A.
Device
IBA Proton Therapy System - PROTEUS 235
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
FDA notice Z-1497-2026