Company record
Lutronic Corporation
0 FDA warning letters since 2023 and 2 recalls in the last 12 months.
Get an email the day FDA posts anything new about Lutronic Corporation.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 16 | Class II | Lutronic Corporation Device |
Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured… Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn. |
| Mar 18 | Class II | Lutronic Corporation Device |
CLARITY II Laser System; Model No. 1110200210. Reports of devices sparking/popping and potentially burning patients. |