Medtronic, Inc.
1 FDA warning letter since 2023 and 7 recalls in the last 12 months.
Get an email the day FDA posts anything new about Medtronic, Inc..
Warning letters
-
DevicesMedtronic, Inc.
CGMP/QSR/Medical Devices/Adulterated
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Nov 4 | Class II | Medtronic, Inc. Device |
OmniaSecure Model 3930M, LEAD 3930M74 74 CM DF4 MR 3930M74 LEAD 3930M79 79 CM DF4 MR 3930M79 The Medtronic OmniaSecure" Model 3930M lead is a steroid eluting, integrated bipolar, nonretractable screw-in, cath… Affected leads require additional engineering evaluation due to complaints regarding difficulty in achieving fixation during implantation. Potential harm of tissue trauma due to implant difficulty. |
| Nov 4 | Class II | Medtronic, Inc. Device |
CareLink SmartSync Programmer, Model 24970A, Pulse generator, permanent, implantable Software problem has the potential for an inaccurate displayed longevity estimate. |
| Nov 4 | Class II | Medtronic, Inc. Device |
CareLink SmartSync, RF Head Patient Connector, Model 24967, Pulse generator, permanent, implantable Software problem has the potential for an inaccurate displayed longevity estimate. |
| Nov 4 | Class II | Medtronic, Inc. Device |
CareLink SmartSync Device Manager App, Model M01A02, Pulse generator, permanent, implantable Software problem has the potential for an inaccurate displayed longevity estimate. |
| May 6 | Class II | Medtronic, Inc. Device |
Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter. Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available. |
| Jan 21 | Class II | Medtronic Inc Device |
Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090 Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration. |
| Dec 10, 2025 | Class II | Medtronic, Inc. Device |
Aurora EV-ICD and Clinical EV-ICD, single chamber, extravascular implantable cardioverter defibrillators (ICD), Product Number DVEA3E4 There is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur. |