FDA Brief

Company record

MUNEVO GMBH

0 FDA warning letters since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about MUNEVO GMBH.

Recalls

ReportedClassFirmProduct and reason
Aug 12 Class II MUNEVO GMBH
Device
munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System
A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.