FDA Brief

Company record

Oculus Technologies of Mexico, S.A. de C.V.

0 FDA warning letters since 2023 and 1 recall in the last 12 months.

Get an email the day FDA posts anything new about Oculus Technologies of Mexico, S.A. de C.V..

Recalls

ReportedClassFirmProduct and reason
Jun 10 Class II Oculus Technologies of Mexico, S.A. de C.V.
Device
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.