0 FDA warning letters since 2023 and 2 recalls in the last 12 months.

Get an email the day FDA posts anything new about Permobil Inc.

Recalls

ReportedClassFirmProduct and reason
Aug 5 Class II Permobil Inc
Device
Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3
This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
Aug 5 Class II Permobil Inc
Device
M5 Corpus Power Wheelchair REF,120032
This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.