Recalls

ReportedClassFirmProduct and reason
Sep 30 Class II SEASPINE ORTHOPEDICS CORPORATION
Device
SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12
The proximal thread depth of the spinal device is out of design specification.
Jul 29 Class II SEASPINE ORTHOPEDICS CORPORATION
Device
SeaSpine Polyaxial Head REF Number: 41-3010
Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.
Dec 24, 2025 Class II SEASPINE ORTHOPEDICS CORPORATION
Device
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0…
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.