FDA Brief

Company record

Securitas Healthcare LLC

0 FDA warning letters since 2023 and 2 recalls in the last 12 months.

Get an email the day FDA posts anything new about Securitas Healthcare LLC.

Recalls

ReportedClassFirmProduct and reason
Jan 14 Class II Securitas Healthcare LLC
Device
Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
FDA notice Z-1032-2026
Jan 14 Class II Securitas Healthcare LLC
Device
Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
FDA notice Z-1031-2026