Recalls

ReportedClassFirmProduct and reason
Aug 19 Class II Spacelabs Healthcare, Inc.
Device
Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers r…
Due to off-the-shelf display port component that has audio/video failures
May 13 Class II Spacelabs Healthcare, Inc.
Device
Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496 Software Version: N/A Product Description: The Command Module (91496) is the core of the Spacelabs patient…
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Apr 15 Class II Spacelabs Healthcare, Inc.
Device
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.