Company record
Steris Corporation
0 FDA warning letters since 2023 and 16 recalls in the last 12 months.
Get an email the day FDA posts anything new about Steris Corporation.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Oct 21 | Class II | Steris Corporation Device |
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115 If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity. |
| Oct 21 | Class II | Steris Corporation Device |
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117 If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity. |
| Oct 21 | Class II | Steris Corporation Device |
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116 If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity. |
| Oct 14 | Class II | Steris Corporation Device |
VERIFY Biological Steam Test Pack Model/Catalog Number: S3065 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction |
| Oct 14 | Class II | Steris Corporation Device |
VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction |
| Oct 14 | Class II | Steris Corporation Device |
VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction |
| Oct 14 | Class II | Steris Corporation Device |
VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction |
| Oct 14 | Class II | Steris Corporation Device |
VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction |
| Oct 14 | Class II | Steris Corporation Device |
HexaVue Focus Integration System, a single, all-in-one system that handles operating room video routing, surgical video integration, and environmental control, Software Version: 2.0.18 A software issue may result in unintended mismatch of patient information on manually printed images. Incorrect patient information appearing on printed images could create a documentation discrepancy or incorrect association of clinical images with a patient record. |
| Oct 14 | Class II | Steris Corporation Device |
VERIFY Biological Steam Test Pack Model/Catalog Number: S3069 Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction |
| Jun 24 | Class II | Steris Corporation Device |
HarmonyAIR A-Series Surgical Lighting System The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach. |
| Jun 24 | Class II | Steris Corporation Device |
HarmonyAIR A-Series Surgical Lighting System There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads. |
| Jun 24 | Class II | Steris Corporation Device |
ALYON Surgical Lighting System The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach. |
| Feb 18 | Class II | Steris Corporation Device |
Barco MNA with the HexaVue IP Integration System, Model/Catalog Number: MNA-6x0-H This is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used. |
| Dec 17, 2025 | Class II | Steris Corporation Device |
AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple… Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users). |
| Dec 17, 2025 | Class II | Steris Corporation Device |
AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simpl… Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users). |