FDA Brief

Company record

Straumann USA LLC

0 FDA warning letters since 2023 and 9 recalls in the last 12 months.

Get an email the day FDA posts anything new about Straumann USA LLC.

Recalls

ReportedClassFirmProduct and reason
Sep 23 Class II Straumann USA LLC
Device
GM Helix Implant, REF: 109.984;
There is a discrepancy between the product identification on the packaging and the dimensions of the implant contained inside.
Aug 19 Class II Straumann USA LLC
Device
Straumann Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim (Article 021.5314); implant, endosseous, root-form
Bone Level Tapered (BLT) Implant, D 4.1mm RC, SLActive 14mm, Roxolid, Loxim may inadvertently contain implants that are 12mm in length instead of 14mm as the labeling would suggest.
Aug 19 Class II Straumann USA LLC
Device
GM Helix Implant, Endosseous Dental Implant Article No. 109.984
There may be a mismatch between the implant contained inside the packaging and the information described on the outer carton label. If the the incorrect implant is installed, the issue could cause bone compression or difficulty during installation in situations where there is sufficient bone volume
May 6 Class II Straumann USA LLC
Device
Straumann n!ce Zr, HT, Full-arch Restoration. Article 010.0158. Endosseous Dental Implant Abutments
Includes an incorrect screw seat interface.
May 6 Class II Straumann USA LLC
Device
Straumann n!ce PMMA Full-arch Restoration. Screw-retained Bridge. Article 010.0304. Endosseous Dental Implant Abutments
Includes an incorrect screw seat interface.
Apr 1 Class II Straumann USA LLC
Device
Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
Apr 1 Class II Straumann USA LLC
Device
Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;
A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.
Feb 18 Class III Straumann USA LLC
Device
WB Impression Post Closed Tray incl. guide screw, L 13mm, TAN/POM
The mix up of the impression caps provided in the package. The provided impression caps are magenta instead of brown.
Dec 3, 2025 Class II Straumann USA LLC
Device
Emdogain; Article Numbers: (1) 440.230V10, (2) 475.132V10;
The possibility for this product that is intended for demonstration purposes only to be placed in patients.