FDA Brief

Company record

Tandem Diabetes Care, Inc.

0 FDA warning letters since 2023 and 3 recalls in the last 12 months.

Get an email the day FDA posts anything new about Tandem Diabetes Care, Inc..

Recalls

ReportedClassFirmProduct and reason
Nov 11 Class I Tandem Diabetes Care, Inc.
Device
Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078
Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
Oct 7 Class II Tandem Diabetes Care, Inc.
Device
Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
Apr 1 Class II Tandem Diabetes Care, Inc.
Device
Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion …
The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.