Company record
THERAKOS DEVELOPMENT LIMITED
0 FDA warning letters since 2023 and 1 recall in the last 12 months.
Get an email the day FDA posts anything new about THERAKOS DEVELOPMENT LIMITED.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Apr 22 | Class II | THERAKOS DEVELOPMENT LIMITED Device |
Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis… Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss. FDA notice Z-1822-2026 |