0 FDA warning letters since 2023 and 17 recalls in the last 12 months.
Get an email the day FDA posts anything new about Ventec Life Systems, Inc..
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Oct 28 | Class I | Ventec Life Systems, Inc. Device |
VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support. Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy. |
| Sep 30 | Class II | Ventec Life Systems, Inc. Device |
REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve Labeling; Oxygen Cylinders do not contain required medical gas labeling |
| Sep 30 | Class II | Ventec Life Systems, Inc. Device |
REF:HF2PC9 HomeFill Oxygen System, HomeFill Cylinder M9 with Integrated Conserver Labeling; Oxygen Cylinders do not contain required medical gas labeling |
| Sep 30 | Class II | Ventec Life Systems, Inc. Device |
HF2P0ST9 HomeFill Oxygen System HomeFill Cylinder M9 with CGA870 Compatible Post Valve Labeling; Oxygen Cylinders do not contain required medical gas labeling |
| Jul 1 | Class II | Ventec Life Systems, Inc. Device |
VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001 Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition. |
| May 13 | Class I | Ventec Life Systems, Inc. Device |
VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000 Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk |