FDA Brief

Recalls

When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.

5,413 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
Renew PeriProtect, Zinc Oxide 12%, Dimethicone 1%, 113g (4 oz.) per tube, DermaRite Industries LLC, 7777 West Side Avenue, North Bergen, NJ 07047. NDC 61924-435-04
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0174-2026
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
PeriGuard Ointment, Petrolatum 49.9%, packaged in a) 5g (NDC 61924-205-05), b) 3.5oz (NDC 61924-205-04), and c) 7oz (NDC 61924-205-07) tubes, DermaRite Industries LLC, 7777 West Side Avenue, North Bergen, NJ 07047.
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0173-2026
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
DermaMed Ointment, Skin Protectant, Aluminum Hydroxide Gel 2%, Net Wt. 106 g (3.75 oz.) per tube, DermaRite Industries LLC, 7777 West Side Avenue, North Bergen, NJ 07047. NDC 61924-214-04
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0164-2026
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
PeriGiene, Health Care Antiseptic, Chloroxylenol 0.5%, 222mL (7.5 fl. oz.) bottle, DermaRite Industries LLC, 7777 West Side Avenue, North Bergen, NJ 07047. NDC 61924-198-08
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0172-2026
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
DermaSarra, Camphor 0.5%, External Analgesic, 222 mL (7.5 fl. oz.) per bottle, DermaRite Industries LLC, 7777 West Side Avenue, North Bergen, NJ 07047. NDC 61924-189-08
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0165-2026
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
DermaFungal, Antifungal Cream with 2% Miconazole Nitrate, packaged in a) 5 g (0.18 oz) per packet (NDC 61924-234-05), and b)106 g (3.75 oz.) per tube (NDC 61924-234-04) DermaRite Industries LLC, 7777 West Side Avenue, No…
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0162-2026
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
KleenFoam Antimicrobial Foam Soap, Chloroxylenol 0.5%, 34 FL OZ (1000 mL) per cartridge, DermaRite Industries LLC, 7777 West Side Avenue, North Bergen, NJ 07047. NDC 61924-093-34
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0168-2026
Nov 26, 2025 Class II DermaRite Industries, LLC
Drug
Lantiseptic, Skin Protectant, Lanolin USP 30%, packaged in a) 5g (0.17 oz) (NDC 61924-304-05), b)14.2g (0.5 oz)( NDC 61924-304-14), c) 4oz (NDC 61924-304-04), d) 4.5oz (NDC 61924-304-45), and e) 12oz (NDC 61924-304-12) t…
CGMP Deviation; manufactured at the same facility where other products and lots were found to be contaminated.
FDA notice D-0170-2026
Nov 26, 2025 Class II Exactech, Inc.
Device
Brand Name: Equinoxe Ergo Product Name: Adjustable Angle rTSA Reamer, 42mm Model/Catalog Number: 321-59-42 Software Version: N/A Product Description: rTSA Reamer, 42mm Component: Not a component
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
FDA notice Z-0535-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202R; 2) ACDF CDS, REF CDS941061M; 3) NEURO PACK-LF, REF DYNJ0038440C; 4) NEURO VP SHUNT/VAGAL NERVE-RF, REF DYNJ21925S; 5) NEURO-ORTHO EXTRA…
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0522-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) HEAD AND NECK, REF CDS984259O; 2) HEAD AND NECK, REF CDS984259P; 3) MINOR ENT HARPER PACK-LF, REF DYNJ0160700F; 4) MAJOR ENT HARPER PACK-LF, REF DYNJ0161710I; 5) THORACO…
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0516-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) GYN LAPAROTOMY HARPER PACK-LF, REF DYNJ0160684F; 2) MAJOR LITHOTOMY PACK-LF, REF DYNJ0425778O; 3) MINOR VAGINAL #76-RF, REF DYNJ27434R; 4) ABDOMINAL HYST PACK, REF DYNJ448…
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0527-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 150 ¿L; Prepared with Balanced Heparin Model 06 0186 Product Description: A clear, plastic capillary blood collection tube coated with Heparin
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0557-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L; Prepared with Lithium Heparin; Green Model/Catalog Number: 07 7250 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device …
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0551-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0518-2026
Nov 26, 2025 Class II Merit Medical Systems, Inc.
Device
PhD Hemostasis Valve, REF: IN4802/K, IN4802/T, IN4918/A, IN7802/C, IN8802/C, IN9802, IN9802/A, K05-02166B, K05T-02138A, K12T-10798A, MAP800/B, MAP800/T, MAP801/B, MAP802/B, MAP802/CNB, MAP802/T, MAP804/B, MAP852/B, SKY18…
Hemostasis valve has a manufacturing defect and when used the valve cap may remain depressed, leading to potential leakage, which may result in procedure inconvenience and/or hemorrhage.
FDA notice Z-0533-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
AFE-T-FILL Micro Capillary Blood Collection, 200 ¿L; Prepared with Lithium Heparin and Plasma Separator Gel; Green Model/Catalog Number: 07 7220 Product Description: The SAFE-T-FILL Capillary Blood Collection System…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0549-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) BREAST BIOPSY PACK-LF, REF DYNJ0160598C; 2) BIOPSY BREAST PACK, REF DYNJ65361A; 3) BREAST BIOPSY/MASTECTOMY, REF DYNJ908839A.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0515-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Collection Tube Extender, Model 09 1004
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0567-2026
Nov 26, 2025 Class II Sterilmed, Inc.
Device
Brand Name: Sterilmed Reprocessed Imaging Catheter Product Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135936 Software Version: N/A Product Descripti…
Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.
FDA notice Z-0574-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE STERILE BASIC EYE PACK 16: 1) REF: DYNJ37743D; 2) REF: DYNJ37743F.
Medline has identified specific sterile eye procedure kits with pouch packaging that may exhibit open seals.
FDA notice Z-0510-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) MINOR SINGLE BASIN PACK, REF DYNJ37698L; 2) BASIN PACK, REF DYNJ50123D; 3) MAJOR BASIN PACK, REF DYNJ59816A; 4) OSC BASIN PACK, REF DYNJ68217A.
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0520-2026
Nov 26, 2025 Class II ELEKTA SOLUTIONS AB
Device
MOSAIQ Oncology Information System Software Builds 3.1.3, 3.2.1, 3.2.2 & 3.2.3.0 Software Version and UDI Codes: 3.1.3.0 (01)07340201500026(10)3.1.3.0 3.1.3.1 (01)07340201500026(10)3.1.3.1 3.1.3.2 (01)07340201500…
When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.
FDA notice Z-0503-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) GENERAL LAPAROSCOPY, REF CDS930027W; 2) GENERAL MAJOR CDS, REF CDS980245X; 3) BASIC GENERAL CDS, REF CDS980566F; 4) TOTAL KNEE CDS, REF CDS982823R; 5) BASIC CDS, REF CDS…
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0521-2026
Nov 26, 2025 Class II Exactech, Inc.
Device
Brand Name: Equinoxe Ergo Product Name: Ergo Augments Pilot Tip Converter Model/Catalog Number: 301-11-11 Software Version: N/A Product Description: Pilot Tip Converter Piece Component: Not a component
Due to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
FDA notice Z-0537-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 115 ¿L; Prepared with Sodium Heparin Model/Catalog Number: 06 0005 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0560-2026
Nov 26, 2025 Class II Thoratec LLC
Device
Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, HeartMate 3 LVAD KIT, OUS; 106524 Software Version: N/A Product…
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
FDA notice Z-0505-2026
Nov 26, 2025 Class II Auris Health, Inc
Device
MONARCH Bronchoscope. Model Number: MBR-000211-B
Potential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks.
FDA notice Z-0543-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) NEURO CRANIOTOMY CDS #36-RF, REF CDS780119V; 2) BASIC BACK CDS, REF CDS780147N; 3) INSTRUMENTED BACK CDS, REF CDS780148R; 4) NEURO-LAMI CDS, REF CDS780208I; 5) TOTAL HIP…
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0530-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) GENERAL AAA #11-RF, REF CDS840261AB; 2) CAROTID ENDARTERECTOMY CDS-LF, REF CDS983637I; 3) PERIPHERAL VASCULAR CDS, REF CDS983723J; 4) ACH OPEN HEART, REF CDS983773J; 5) …
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0513-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 115 ¿L; Prepared with Balanced Heparin Model/Catalog Number: 06 0185 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0563-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L; Prepared with Clot Activator and Serum Separator Gel; Red Model/Catalog Number: 07 7121 Product Description: The SAFE-T-FILL Capillary Blood Collection System is…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0546-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) MAJOR LAPAROTOMY CDS, REF CDS860015W; 2) MAJOR LAPAROTOMY CDS, REF CDS860015X; 3) MINOR LAPAROTOMY CDS, REF CDS860016R; 4) MINOR LAPAROTOMY CDS, REF CDS860016S; 5) LAPAR…
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0517-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as LOCAL BASIC, REF DYNJ34418M
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0511-2026
Nov 26, 2025 Class II Sterilmed, Inc.
Device
Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Product Name: CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135910 S…
Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.
FDA notice Z-0573-2026
Nov 26, 2025 Class II Thoratec LLC
Device
Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II¿ LVAS Kit, with Pocket Controller, Seale…
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
FDA notice Z-0507-2026
Nov 26, 2025 Class II Beckman Coulter Inc.
Device
DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping up to ten fluorescent detection channels us…
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
FDA notice Z-0541-2026
Nov 26, 2025 Class II Berkeley Advanced Biomaterials, LLC
Device
Tribio Implant 7.5mm x 40mm REF M80 SB008 UDI code: (01) 00816125023653(10) EBL10B13C(17) 270630 Tribio" Implant is indicated for use in bony voids or gaps of the skeletal system (i.e. extremities, pelvis and postero…
Due to incorrect product label (Incorrect product name identified on outer packaging).
FDA notice Z-0540-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L; Prepared with Clot Activator; Amber Microtube, Red Model/Catalog Number: 07 7450 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a coll…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0554-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Magnetic Rings for SAFE-T-FILL Blood Gas Capillaries, Model 07 9502
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0565-2026
Nov 26, 2025 Class II Becton Dickinson & Company
Device
BD Luer Tip Caps; Catalog Number(s) or Model Number(s): 308341;
BD has confirmed through internal testing that certain device trays failed routine biocompatibility testing.
FDA notice Z-0504-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: CATH LAB PACEMAKER PACK, REF DYNJ59818A
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0512-2026
Nov 26, 2025 Class II Quest International, Inc.
Device
Brand Name: ReQuest Measles IgG Product Name: Rubeola IgG Enzyme Linked Immunoabsorbant Assay Model/Catalog Number: 01-190 Software Version: N/A Product Description: Measles IgG tests, performed using ELISA (Enzyme-L…
Due to distributing the measles IgG IVD without a premarket approved/cleared.
FDA notice Z-0534-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL MicroHematocrit Capillary Tubes, 50 ¿L; Plastic, Ammonium Heparinized, Red Marking Model/Catalog Number: 06 0507 Product Description: A clear, plastic capillary blood collection tube coated with Heparin
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0555-2026
Nov 26, 2025 Class II Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) C-SECTION OR 30 CDS, REF CDS820075W; 2) C-SECTION, REF CDS984261J; 3) C-SECTION CDS, REF CDS985282F; 4) C-SECTION PACK, REF DYNJ02577D; 5) C-SECTION PACK-LF, REF DYNJ053…
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
FDA notice Z-0526-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L; Prepared with Clot Activator; Red Model/Catalog Number: 07 7151 Product Description: The SAFE-T-FILL Capillary Blood Collection System is a collection device use…
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0548-2026
Nov 26, 2025 Class II Thoratec LLC
Device
Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Con…
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
FDA notice Z-0508-2026
Nov 26, 2025 Class I TMJ Solutions Inc
Device
TMJ Bilateral Implant REF: CHG020 Sterile EO, Rx Only
Patient-fitted temporomandibular joint implant contained discrepancy in screw hole position.
FDA notice Z-0488-2026
Nov 26, 2025 Class II Greiner Bio-One North America, Inc.
Device
Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard S…
Some of the Vacuette¿ CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters.
FDA notice Z-0571-2026
Nov 26, 2025 Class II ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Device
SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 150 ¿L; Untreated. Model 06 0199 Product Description: A clear, plastic capillary blood collection tube
Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.
FDA notice Z-0559-2026