When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.

5,412 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Sep 23 Class II PTW FREIBURG
Device
Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
Sep 23 Class II NOXBOX LTD
Device
NOxBOXi Nitric Oxide Delivery System.
Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
Sep 23 Class II ICU Medical, Inc.
Device
IV Sets with drip chambers
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Sep 23 Class II Medline Industries, LP
Device
Convenience kits MYRINGOTOMY PACK-LF DYNJ59097A
Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.
Sep 23 Class I AVID Medical, Inc.
Device
Haylard BASIC BIOPSY TRAY; Model/Item Number (Kit): PM415;
Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.
Sep 23 Class II Maquet Cardiopulmonary Gmbh
Device
Percutaneous Insertion Kit (PIK), Guidewire 150 cm, Model Numbers: 701053069 (USA), 701047385 (OUS);
The packaging may not consistently maintain sterile barrier integrity.
Sep 23 Class I Abiomed, Inc.
Device
Impella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-00…
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Sep 23 Class II Boston Scientific Corporation
Device
Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300…
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
Sep 23 Class I Medline Industries, LP
Device
Medline Convenience Kits: 1) DYNDB3086D, Catalog Number: DYNDB3086D; 2) DYNDB3100, Catalog Number: DYNDB3100; 3) DYNDH1619A, Catalog Number: DYNDH1619A; 4) DYNDH1620A, Catalog Number: DYNDH1620A; 5) DYNDH1694A, Catal…
The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
Sep 23 Class I Medline Industries, LP
Device
Medline Convenience Kits: 1) OPEN HEART ADULT, Catalog Number: CDS840396Y; 2) FULL LINE ANEURYSM CDS, Catalog Number: CDS983672O; 3) NON STERILE CHEST TUBE REMOVAL, Catalog Number: DYNDA1080
The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
Sep 23 Class II GE Healthcare
Device
Centricity Enterprise Archive V2, V3, and V4, software product for receiving, archiving, and sending electronic medical data, All Versions
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Sep 23 Class II ICU Medical, Inc.
Device
IV Non-Dedicated Gravity Administration Sets with drip chambers
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
Sep 23 Class II Howmedica Osteonics Corp.
Device
TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Sep 23 Class II Howmedica Osteonics Corp.
Device
OMNIFIT SERIES I ACET. INSERT 10¿ 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Sep 23 Class I Medline Industries, LP
Device
Medline Convenience Kits: 1) DYNDH1193, Catalog Number: DYNDH1193; 2) DYNDH1487A, Catalog Number: DYNDH1487A
The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
Sep 23 Class I Abiomed, Inc.
Device
Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 2. Impella Optical Controller, Packaged, EU; Produc…
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Sep 23 Class II Maquet Cardiopulmonary Gmbh
Device
Percutaneous Insertion Kit (PIK), Guidewire 100 cm, Model Numbers: 701053068 (USA), 701047384 (OUS);
The packaging may not consistently maintain sterile barrier integrity.
Sep 23 Class II Wilson-Cook Medical Inc.
Device
Brand Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Product Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Model/Catalog Number: HEMO-7, HEMO-10, HEMO-7-EU, HEMO-10-EU Software Version: N/A Product Description: Hemostatic device for endo…
Due to issue with endoscope adherence and inability to spray powder. Firm will provide an updated Instruction for Use (IFU).
Sep 23 Class II Bayer Medical Care, Inc.
Device
MEDRAD¿ Stellant FLEX Syringe Kit used with Stellant FLEX CT Injection System; Catalog number: FLEXS-150-QFT;
Some devices may not have been properly sterilized due to a manufacturing sterilization issue.
Sep 23 Class II PI CARDIA LTD
Device
Pi-Cardia Inc. ShortCut REF: Z-A-20.003-US
Transfemoral catheter, positing arm may detached due to weld issue.
Sep 23 Class II GE Healthcare
Device
Centricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and processing of digital images and data acquired from…
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Sep 23 Class II Stryker Sustainability Solutions
Device
REF:B5LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X100MM
Reusable access port system may crack or seperate during use.
Sep 23 Class II Howmedica Osteonics Corp.
Device
ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INSERT 48MM 69-2248 ALL POLY CONSTRAINED INSERT 50MM 69-…
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Sep 23 Class II Howmedica Osteonics Corp.
Device
ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID C-M071-6-700 ARC2F ACETABULAR INSERT 50mm OD x 28mm ID 10DEGREE HOOD C-M071-8-400 …
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Sep 23 Class II Mirion Technologies (Capintec), Inc.
Device
Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;
Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
Sep 23 Class II Stryker Sustainability Solutions
Device
REF:B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X75MM
Reusable access port system may crack or seperate during use.
Sep 23 Class II F & A Foundation
Device
Reign Medical Clench Staple System (Compression), Description/REF: 9mm x 7mm/152-30907; 11mm x 10mm/152-31110; 15mm x 12mm/152-31512; 18mm x 18mm/152-31818; 20mm x 20mm/152-32020; 9mm x 7mm/152-40907; 11mm x 10m…
Bone fixation staple requires a new 510(k); the performance characteristics of this device have not been adequately established.
Sep 23 Class II Howmedica Osteonics Corp.
Device
SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Sep 23 Class I Medline Industries, LP
Device
Medline Convenience Kits: 1) KIT ANESTHESIA L&D A- LINE SET, Catalog Number: DYKMBNDL199A
The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
Sep 23 Class II Philips North America
Device
Philips PIC iX Essentials REF:867093
Due to software malfunction, patient information system may freeze and become unresponsive.
Sep 23 Class II ICU Medical, Inc.
Device
IV Sets with drip chambers
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Sep 23 Class II ICU Medical, Inc.
Device
Plum administration sets
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Sep 23 Class I Nova Biomedical Corporation
Device
StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
Sep 23 Class I Nova Biomedical Corporation
Device
StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
Sep 23 Class I AVID Medical, Inc.
Device
Haylard UNIV IMAGING BIOPSY TRUCUSTOM; Model/Item Number (Kit): VARV042-03;
Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.
Sep 23 Class I Abiomed, Inc.
Device
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: …
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Sep 23 Class II GE Healthcare
Device
Centricity PACS Server V7, intended for managing digital medical images and associated data, All versions
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Sep 23 Class II Happiest Baby, Inc
Device
Happiest Baby, Inc SNOO Sleep Sack, X-Small Model/Catalog Number: S2101-101 S2101-108-6 S2101-133 S2102-112 S2107-101 S2107-110
Extra small (ES) and Extra-large (XL) infant sleep sacks have not been reviewed for safety and effectiveness by the FDA.
Sep 23 Class II ICU Medical, Inc.
Device
IV Sets with drip chambers
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Sep 23 Class II Medline Industries, LP
Device
Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N
The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.
Sep 23 Class II ICU Medical, Inc.
Device
Oncology sets with drip chambers
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
Sep 23 Class II Datex-Ohmeda, Inc.
Device
Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Sep 23 Class II ICU Medical, Inc.
Device
Oncology sets with drip chambers
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
Sep 23 Class I Nova Biomedical Corporation
Device
StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.
Sep 23 Class II GE Medical Systems Information Technologies Inc
Device
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
Sep 23 Class I Medline Industries, LP
Device
Medline Convenience Kits: 1) DYKM1000B, Catalog Number: DYKM1000B; 2) DYKM2187A, Catalog Number: DYKM2187A
The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
Sep 23 Class I AVID Medical, Inc.
Device
Haylard 18G SINGLE SHOT EPI TRAY; Model/Item Number (Kit): DPGE002-02;
Potential for product quality and sterility issues regarding kits containing Sodium Chloride ampules subject to recall by Spectra Medical.
Sep 23 Class II Datex-Ohmeda, Inc.
Device
Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Sep 23 Class II Datex-Ohmeda, Inc.
Device
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Sep 23 Class II Stryker Sustainability Solutions
Device
REF:2B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/OPTIVIEW TECHNOLOGY & SLEEVE 5MM X 75MM
Reusable access port system may crack or seperate during use.