ReportedClassFirmProduct and reason
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 50 mcg (0.050 mg), 10 Tablets (10 x 1) unit dose blisters per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. Distributed by Cardinal He…
Subpotent Drug
Jul 29 Class II Precision Dose Inc.
Drug
Glycopyrrolate Oral Solution, 1 mg/5 mL, delivers 5 mL, Rx only, Pkg: Precision Dose, Inc, S. Beloil, IL 61080, unit-dose oral syringes, NDCs 68094-073-01 (oral syringe) and 68094-073-58 (case).
Failed Impurities/Degradation Specifications
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 125 mcg (0.0125 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6955-61
Subpotent Drug
Jul 29 Class II Fresenius Kabi USA, LLC
Drug
Glycopyrrolate Injection, USP, 1 mg per 5 mL (0.2 mg per mL), 25 x 5 mL fill in a 6.5 mL vial, Rx only, Multiple Dose Vials, Fresenius Kabi, Lake Zurich, IL 60047 NDC Tray: 63323-578-05; NDC Vial: 63323-578-07
Presence of Particulate Matter: Hair was found in product
Jul 29 Class II Fresenius Kabi USA, LLC
Drug
Xylocaine (lidocaine HCl Injection, USP), 1%, 500 mg per 50 mL (10 mg per mL), 25 Multiple-Dose Vials, 50 mL, Rx only, Sterile, Fresenius Kabi, Lake Zurich, IL 60047 NDC Tray: 63323-485-57; NDC Vial: 63323-485-03
Presence of Particulate Matter: Hair was found in products
Jul 29 Class II Rohto-Mentholatum (Vietnam) Co., Ltd.
Drug
Rohto Cooling Eye Drops Optic Glow (naphazoline hydrochloride 0.03%, povidone 0.5%, propylene glycol 0.2%), Sterile, 0.4 FL OZ (13 mL), Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UP…
Lack of Assurance of Sterility
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 112 mcg (0.0112 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6954-61
Subpotent Drug
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 25 mcg (0.025 mg), 10 Tablets (10 x 1) unit dose blisters per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. Distributed by Cardinal He…
Subpotent Drug
Jul 29 Class II Ascend Laboratories, LLC
Drug
Amlodipine and Olmesartan Medoxomil Tablets, 10 mg/20 mg, 30 Tablets per bottle, Rx Only, Manufactured by: Alkem Laboratories Ltd., INDIA; Distributed by: Ascend Laboratories, LLC., Parsnippany, NJ 07054. NDC: 67877-500…
Failed Dissolution Specifications; Olmesartan Medoxomil content below specifications
Jul 29 Class III Asclemed USA Inc.
Drug
Lidolog Kit, Kit Contains: Lidocaine HCl Injection, USP, 2% (2mL) vial, 1 Dose, Single use Only, Rx Only, Distributed by: Enovachem Pharmaceuticals, Torrance, CA 90501, NDC: 76420-760-01.
Labeling: Not Elsewhere Classified: The label wrap covers the barcode, making it hard to scan
Jul 29 Class II Bionpharma Inc.
Drug
Divalproex Sodium Delayed-Release Tablets, USP, 500 mg, 500 Tablets bottles, Rx only, Distributed by: Bionpharma Inc., Princeton, NJ 08540, Made in India, NDC 69452-435-30.
Presence of Foreign Tablets/Capsules
Jul 29 Class II Rohto-Mentholatum (Vietnam) Co., Ltd.
Drug
Rohto Cooling Eye Drops Max Strength (naphazoline hydrochloride 0.03%, polysorbate 80 0.2%), Sterile, TWIN PACK, 0.4 FL OZ (13 mL), Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UPC 31…
Lack of Assurance of Sterility
Jul 29 Class II Fresenius Kabi USA, LLC
Drug
Lidocaine HCl Injection, USP, 2%, 100 mg per 5 mL (20 mg per mL), 25 Single Dose Vials, 5 mL per vial, Rx only, Sterile, Fresenius Kabi, Lake Zurich, IL 60047 NDC Tray: 63323-208-05; NDC Vial: 63323-208-01
Presence of Particulate Matter: Hair was found in products
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 50 mcg (0.050 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6950-61
Subpotent Drug
Jul 29 Class III Ascend Laboratories, LLC
Drug
Dabigatran etexilate Capsules, 150 mg, 20-count carton (2x10), Rx only, Manufactured by: Alkem Laboratories, Mumbai, INDIA, Distributed by: Major Pharmaceuticals, Indianapolis, IN 46268, NDC 0904-7255-10.
Failed Impurities/Degradation Specifications: OOS result was observed during the related substance test analysis with respect to Impurity II, Impurity III and Total impurity testing for 12 Month Long-Term Stability.
Jul 29 Class II Rohto-Mentholatum (Vietnam) Co., Ltd.
Drug
Rohto Cooling Eye Drops Dry Aid (povidone 0.68%, propylene glycol 0.3%), Sterile, 0.34 FL OZ (10 mL), Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UPC 310742011135, NDC 10742-8162-1
Lack of Assurance of Sterility
Jul 29 Class II Fresenius Kabi USA, LLC
Drug
Xylocaine (lidocaine HCl Injection, USP), 1%, 200 mg per 20 mL (10 mg per mL), 25 Multiple-Dose Vials, 20 mL, Rx only, Sterile, Fresenius Kabi, Lake Zurich, IL 60047 NDC Tray: 63323-485-27; NDC Vial: 63323-485-01
Presence of Particulate Matter: Hair was found in products
Jul 29 Class II Rohto-Mentholatum (Vietnam) Co., Ltd.
Drug
Rohto Cooling Eye Drops Digi Eye (hypromellose 0.35%, tetrahydrozoline HCl 0.05%), Sterile, 0.4 FL OZ (13 mL), Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UPC 310742010602, NDC 10742…
Lack of Assurance of Sterility
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 25 mcg (0.025 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6949-61
Subpotent Drug
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 125 mcg (0.0125 mg), 10 Tablets (10 x 1) unit dose blisters per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. Distributed by Cardinal …
Subpotent Drug
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 75 mcg (0.075 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6951-61
Subpotent Drug
Jul 29 Class II Fresenius Kabi USA, LLC
Drug
Naropin (ropivacaine hydrochloride Injection, USP), 0.5%, 100 mg per 20 mL (5 mg per mL), Twenty-five, 20 mL Single-Dose Vials, Rx only, Fresenius Kabi, Lake Zurich, IL 60047 NDC Tray: 63323-286-23; NDC Vial: 63323-286-…
Presence of Particulate Matter: Hair was found in products
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 88 mcg (0.088 mg), 100 Tablets (10 x 10 unit dose blisters) per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. NDC: 0904-6952-61
Subpotent Drug
Jul 29 Class III Supernus Pharmaceuticals, Inc.
Drug
Trokendi XR, topiramate extended-release capsules, 25mg, 30 Capsules, Rx only, Manufactured by: Catalent Pharma Solutions, Winchester, KY 40391 USA, Manufactured for: Supernus Pharmaceuticals, Inc., Rockville, MD 20850 U…
Failed Dissolution Specifications
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 150 mcg (0.0150 mg), 10 Tablets (10 x 1) unit dose blisters per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. Distributed by Cardinal …
Subpotent Drug
Jul 29 Class II Rohto-Mentholatum (Vietnam) Co., Ltd.
Drug
Rohto Cooling Eye Drops Cool Relief (naphazoline hydrochloride 0.012%, polysorbate 80 0.2%), Sterile, 0.4 FL OZ (13 mL) each, Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UPC 31074201…
Lack of Assurance of Sterility
Jul 29 Class II Major Pharmaceuticals
Drug
Levothyroxine Sodium Tablets, USP, 75 mcg (0.075 mg), 10 Tablets (10 x 1) unit dose blisters per carton, Rx only, Packaged and Distributed by: MAJOR PHARMACEUTICALS, Indianapolis, IN 46268 USA. Distributed by Cardinal He…
Subpotent Drug
Jul 29 Class II Rohto-Mentholatum (Vietnam) Co., Ltd.
Drug
Rohto Cooling Eye Drops Max Strength (naphazoline hydrochloride 0.03%, polysorbate 90, 0.2%), Sterile 0.4 FL OZ (13 mL), Distributed by: The Mentholatum Company, Orchard Park, NY 14127, Made in Vietnam, UPC 310742011012…
Lack of Assurance of Sterility
Jul 29 Class II Olympus Corporation of the Americas
Device
Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.
Potential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
Jul 29 Class I Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) TOTAL JOINT KIT, Medline Kit SKU DYKS1574A; 2) TOTAL KNEE BASIC, Medline Kit SKU DYNJ902378J; 3) TOTAL HIP, Medline Kit SKU DYNJ903492K; 4) ANTERIOR HIP, Medline Kit SKU D…
Medline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in Dextrose Injection, USP. B. Braun Medical Inc. was notified by their supplier, Spectra Medical Devices, LLC of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. Huons Co, Ltd. has issued a voluntary recall of this product due to quality issues.
Jul 29 Class II Boston Scientific Corporation
Device
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Jul 29 Class II Boston Scientific Corporation
Device
FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Jul 29 Class II Medline Industries, LP
Device
Medline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397P
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Jul 29 Class II Prismatik Dentalcraft, Inc.
Device
Brand Name: Silent Nite Product Name: Silent Nite Sleep Appliance Model/Catalog Number: SILNT Product Description: Silent Nite Sleep Appliance is a mandibular advancement device (MAD) intended to treat snoring and …
Due to device not having FDA Clearance
Jul 29 Class II PIE Medical Imaging B.V.
Device
3mensio Workstation (Vascular Fenestrated) software
User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
Jul 29 Class II SEASPINE ORTHOPEDICS CORPORATION
Device
SeaSpine Polyaxial Head REF Number: 41-3010
Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.
Jul 29 Class II Boston Scientific Corporation
Device
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Jul 29 Class II Emano Metrics, Inc.
Device
Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acous…
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
Jul 29 Class II Medline Industries, LP
Device
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. PROSTATE ROBOTIC DYNJ901822B
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Jul 29 Class II Boston Scientific Corporation
Device
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.
Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
Jul 29 Class II Medline Industries, LP
Device
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. SM-LAPAROSCOPIC PROSTATE PAC DYNJ61206J
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Jul 29 Class II Olympus Corporation of the Americas
Device
Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4
Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
Jul 29 Class I Medline Industries, LP
Device
MEDLINE medical procedure convenience kit labeled as: HIP CDS-LF, Medline kit SKU CDS984556N
The procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were recalled due to quality issues that could result in drug ineffectiveness.
Jul 29 Class II Medline Industries, LP
Device
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. RR-DAVINCI PROSTATE ACCESSORY DYNJ63392F
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Jul 29 Class II Micro Therapeutics, Inc.
Device
Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12. Intracranial aneurysm flow diverter
Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.
Jul 29 Class II Medline Industries, LP
Device
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. DAVINCI PROSTATE PACK DYNJ68477L DYNJ68477M
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Jul 29 Class II Medline Industries, LP
Device
Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
Jul 29 Class II Medline Industries, LP
Device
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. NON MB DAVINCI LITHOTOMY SM PK DYNJT6105
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Jul 29 Class I Medline Industries, LP
Device
MEDLINE convenience kits labeled as: 1) C-SECTION, Medline SKU DYNJ910835; 2) C-SECTION, Medline SKU DYNJ910835A; medical convenience kits
Medline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in Dextrose Injection, USP. B. Braun Medical Inc. was notified by their supplier, Spectra Medical Devices, LLC of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. Huons Co, Ltd. has issued a voluntary recall of this product due to quality issues.
Jul 29 Class II KENT IMAGING, INC.
Device
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.