ReportedClassFirmProduct and reason
Oct 28 Class II PRITAM INTERNATIONAL PRIVATE LIMITED
Drug
Brush Buddies, Mint Chocolate Chip Ice Cream Flavored Toothpaste, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-52…
CGMP Deviations.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/STRAIGHT/1RO/130; Material Number: M001195710
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class I Ventec Life Systems, Inc.
Device
VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE DRAPE C-SECTION W/POUCH ST 7/CS, Medline Product Number/SKU DYNJP6101
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II TECHNO-PATH MANUFACTURING LTD.
Device
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/130; Material Number: M001195210
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class I Ventec Life Systems, Inc.
Device
VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/155; Material Number: M001195330
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKU DYNJ16127N; 2) NYGH-DELIVERY PACK-LF, Medline Kit Number/SKU DYNJ27206D; 3) OB/GYN, Medline Kit Number/SKU DYNJ9…
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class I Ventec Life Systems, Inc.
Device
V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class I Ventec Life Systems, Inc.
Device
V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class II Medtronic Neuromodulation
Device
InterStim SureScan MRI Leads 978A1 & 978B1: LEAD 978A128 ISTM MICRO 28CM SSMRI TINED 978A128 LEAD 978A133 ISTM MICRO 33CM SSMRI TINED 978A133 LEAD 978A141 ISTM MICRO 41CM SSMRI TINED 978A141 LEAD 978B128…
Wrong lead or percutaneous extension selected. Use of an incompatible lead or percutaneous extension will prevent the system from working as intended and may result in delay of surgery or need for surgical replacement of the product.
Oct 28 Class II Solstice Corporation
Device
Xact mole marker for mammography Model/Catalog Number: M15RM
The radiopaque material may detach from the marker. In some cases, artifact was noted and attributed to the detached material.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/130; Material Number: M001195320
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/130; Material Number: M001195270
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: 1) FETAL SURGERY CDS-LF, Medline Kit Number/SKUCDS840214R; 2) MAJOR, Medline Kit Number/SKUDYNJ904851I; 3) ABDOMINAL TRANSPLANT, Medline Kit Number/SKUDYNJ908124B.
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II restor3d Inc.
Device
Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use impla…
Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
Oct 28 Class II CooperVision, Inc.
Device
Generic Name: Softmed Breathables Advanced Toric (fanfilcon A) Soft (Hydrophilic) Contact Lens; Model/Catalog Number: Softmed Breathables Advanced Toric Product Description: 55% water content, toric, silicone hydrogel …
The affected products bear an incorrect statement on the carton packaging indicating that the product is "Indicated for Extended Wear". However, this product is intended for daily wear, and the extended-wear statement should not have appeared on the affected carton labeling. This product is not approved by the U.S. Food and Drug Administration (FDA) for extended wear use.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/155; Material Number: M001195650
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/155; Material Number: M001195850
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/155; Material Number: M001195480
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: C-SECTION PACK, Medline Product Number/SKU DYNJ48455A
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/130; Material Number: M001195640
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: TUBAL LIGATION RFD, Medline Product Number/SKU DYNJ906291G
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/105; Material Number: M001195430
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II AVID Medical, Inc.
Device
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31.
A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
Oct 28 Class II Beckman Coulter, Inc.
Device
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Mode…
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
Oct 28 Class II Abbott
Device
Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5. 9-ACP2-010-025 6. 9-ACP2-010-028 7. 9-ACP2-010-031 8. 9-ACP2-010-034
Changes to Instructions for Use (IFU) for the device regarding risk of Pulmonary Artery Injury (PAI).
Oct 28 Class II Medline Industries, LP
Device
MEDLINE DRAPE C-SEC FEN CLEAR SCRN WIRE, Medline Product Number/SKU DYNJP6106
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/155; Material Number: M001195510
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/130; Material Number: M001195470
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II TECHNO-PATH MANUFACTURING LTD.
Device
Techno-Path Multichem AE Unassayed Level 1 REF:AE621X
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Oct 28 Class I Ventec Life Systems, Inc.
Device
VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/2RO/130; Material Number: M001195660
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/2RO/130; Material Number: M001195670
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class I Philips Ultrasound, LLC
Device
Philips Lumify Diagnostic Ultrasound System
Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/130; Material Number: M001195610
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class I Ventec Life Systems, Inc.
Device
VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class II Medtronic Navigation, Inc.
Device
Medtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly
Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/105; Material Number: M001195400
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class I Ventec Life Systems, Inc.
Device
VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: C-SECTION TRAY-LF, Medline Kit Number/SKU OBN471K
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/STRAIGHT/1RO/155; Material Number: M001195720
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Oct 28 Class II AVID Medical, Inc.
Device
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30.
A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
Oct 28 Class I Ventec Life Systems, Inc.
Device
VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Oct 28 Class II Philips North America
Device
Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;
Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
Oct 28 Class II Medline Industries, LP
Device
MEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215P
Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Oct 28 Class II Boston Scientific Cork, Ltd.
Device
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/BERN/1RO/130; Material Number: M001195740
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.