  1. Next Medical Products Company, LLC

    FDA warns Next Medical Products that its ultrasound gels were contaminated and made without required quality controls or clearance.

    High risk
  2. Chirp Innovations, LLC

    FDA warned Chirp Innovations over quality-system gaps, missing adverse-event reporting procedures, and unapproved marketing of its Halo muscle stimulator for menstrual pain.

    High risk
  3. Olympus Medical Systems Corporation

    FDA warns Olympus over quality-system failures across six medical device plants, including unvalidated designs and released nonconforming devices.

    High risk
  4. Unilam Co., LTD

    Unilam Co., LTD's UV sunlamps violate quality-system and radiation-safety testing rules, FDA says.

    High risk
  5. Specialty Health Products CO, LLC

    Specialty Health Products was cited for quality-system, unapproved marketing, and listing violations for its Hydro-San and Aqua Cleanse colonic irrigation devices.

    High risk
  6. Nipro Renal Solutions USA, Corporation

    Nipro Renal Solutions USA released hemodialysis dialysate lots with out-of-spec endotoxin (bacterial toxin) and had widespread quality-system failures, FDA says.

    High risk
  7. Koven Technologies, Inc.

    FDA warns Koven Technologies that its Bidop 3 Doppler was labeled for fetal use without clearance and had quality system violations.

    High risk
  8. Linemaster Switch Corporation

    Linemaster Switch Corporation, a Connecticut maker of foot pedals for medical lasers, was cited for quality system failures after an FDA inspection.

    High risk
  9. ZIIP, Inc.

    FDA warns ZIIP over quality-system violations and unapproved marketing of HALO 2.0, DOT, and gels.

    High risk
  10. AseptiKits, LLC

    AseptiKits marketed unapproved sterile-preparation device kits without FDA clearance or approval, FDA says.

    High risk
  11. 3B Medical, Inc. dba Reach Health, Inc.

    FDA warns Reach Health over unapproved APAP use change and complaint/CAPA failures.

    High risk
  12. BMC Medical Co., Ltd.

    BMC Medical’s G3X APAP devices were misbranded/adulterated after significant changes without a new 510(k) and an unreported firmware fix.

    High risk
  13. ZOLL Medical Corporation

    FDA warned ZOLL Medical after an inspection found quality-system, design-validation, supplier, and reporting failures for defibrillators, ventilators, and electrodes.

    High risk
  14. Estar Technologies Ltd

    FDA warns Estar Technologies over selling seven unapproved PRP/PRF device kits in the U.S. without clearance or approval.

    High risk
  15. Unetixs Vascular, Inc.

    Unetixs Vascular marketed modified ROODRA and 2CP vascular diagnostic devices without required FDA clearance.

    High risk
  16. Medline Industries, LP

    Medline's Glen Falls, NY plant made angiographic syringes and kits under flawed CAPA, cleaning and design verification controls, FDA says.

    High risk
  17. Physitemp Instruments, LLC

    FDA says Physitemp sold clinical thermometers and temperature probes without FDA clearance and with unvalidated sterilization and welding.

    High risk
  18. Longhorn Vaccines and Diagnostics LLC

    Longhorn Vaccines and Diagnostics sold its PrimeStore MTM sample transport device without required quality controls or clearance for claims beyond its approved uses.

    High risk
  19. IsoTis OrthoBiologics, Inc.

    IsoTis OrthoBiologics in Irvine, CA failed to properly fix and assess possible particulate contamination in its Accell bone filler devices, FDA says.

    High risk
  20. Avertix Medical, Inc.

    FDA warns Avertix Medical over Guardian implantable cardiac event detector quality violations and late injury reports.

    High risk
  21. Exactech, Inc. dba Advita Ortho

    FDA warned Exactech/Advita Ortho over quality-system failures in Equinoxe shoulder implants found during a July 2025 Sarasota inspection.

    High risk
  22. ExThera Medical Corporation

    ExThera shipped Seraph 100 blood filters for compassionate use without FDA approval and promoted them as approved.

    High risk
  23. Beta Bionics, Inc.

    FDA says Beta Bionics' iLet Bionic Pancreas system had quality-system failures, including ignored complaints of severe hypoglycemia and diabetic ketoacidosis.

    High risk
  24. Flextronics America LLC

    FDA warns Flextronics America, contract maker of FreeStyle Libre 3 glucose sensors in Austin, TX, over an unvalidated assembly step and weak corrective actions.

    High risk
  25. Abbott Diabetes Care, Inc.

    FDA warns Abbott Diabetes Care that FreeStyle Libre 3 glucose monitors lacked design transfer and finished-device accuracy release testing.

    High risk
  26. Unomedical Device S.A. de C.V.

    FDA warns Unomedical over insulin infusion sets with unvalidated leak tests and unreported serious injuries.

    High risk
  27. Meridian Bioscience, Inc

    FDA warns Meridian Bioscience's Magellan Diagnostics over quality system failures and unreported problems with its blood lead test kits.

    High risk
  28. MAST Biosurgery USA, Inc.

    MAST Biosurgery USA never ran promised endotoxin tests on its SurgiWrap FROST surgical mesh and failed to report patient injuries, FDA says.

    High risk
  29. ASP Global, LLC. dba Anatomy Supply Partners, LLC.

    FDA cites ASP Global for quality-system, reporting, and unapproved-change violations involving Safe-T-Fill blood collection devices.

    High risk
  30. Palamur Biosciences Private Limited

    FDA warns Palamur Biosciences over GLP violations in medical device safety testing and inadequate corrective actions.

    High risk
  31. Envoy Medical Inc.

    CGMP/QSR/Medical Devices/Adulterated

  32. Hong Qiangxing Shenzhen Electronics Limited

    CGMP/QSR/Medical Devices/Adulterated

  33. Qianjiang Kingphar Medical Material Co Ltd.

    CGMP/QSR/Medical Devices/Adulterated/Misbranded

  34. Contec Medical Systems Co., Ltd.

    CGMP/QSR/Medical Devices/Adulterated

  35. Royal Philips

    CGMP/QSR/Medical Devices/Adulterated/Misbranded

  36. LEVO AG

    CGMP/QSR/Medical Devices/Adulterated

  37. Technological Medical Advancements LLC

    CGMP/QSR/Medical Devices/Adulterated

  38. Miach Orthopaedics

    CGMP/QSR/Medical Devices/Adulterated

  39. LeMaitre Vascular, Inc.

    CGMP/QSR/Medical Devices/Adulterated

  40. Uscom Kft

    CGMP/QSR/Medical Devices/PMA/Adulterated/Misbranded

  41. Les Encres LLC

    CGMP/QSR/Medical Devices/Adulterated

  42. Spectra Therapy, LLC

    CGMP/QSR/Medical Devices/Adulterated

  43. Mectronic Medicale S.R.L.

    CGMP/QSR/Medical Devices/Adulterated

  44. Visgeneer, Inc

    CGMP/QSR/Medical Devices/Adulterated/Misbranded

  45. Onkos Surgical, Inc.

    CGMP/QSR/Medical Devices/Adulterated

  46. Leltek Inc.

    CGMP/QSR/Medical Devices/Adulterated

  47. Aju Pharm Co., Ltd.

    CGMP/QSR/Medical Devices/Adulterated

  48. Reset Technology Corporation

    CGMP/QSR/Medical Devices/Adulterated

  49. DFI Co., Ltd.

    CGMP/QSR/Medical Devices/Adulterated

  50. Sedecal S.A.

    CGMP/QSR/Medical Devices/Adulterated